EKOM DK50 DE Používateľská príručka

Typ
Používateľská príručka
User manual
EN
Руководство пользователя
RU
Benutzerhandbuch
DE
Návod na použitie
SK
DK50 DE
MEDICAL COMPRESSOR
MEDIZINALKOMPRESSOR
МЕДИЦИНСКИЙ КОМПРЕССОР
MEDICINÁLNY KOMPRESOR
DK50 DE
EKOM spol. s r. o.
Priemyselná 5031/18
SK-921 01 Piešťany
Slovak Republic
tel.: +421 33 7967255
fax: +421 33 7967223
www.ekom.sk
DATE OF LAST REVISION
DATUM DER LETZTEN ÜBERARBEITUNG
ДАТА ПОСЛЕДНЕГО ПЕРЕСМОТРА
DÁTUM POSLEDNEJ REVÍZIE 05/2023
NP-DK50 DE-22_05-2023
112000072-000
CONTENTS
4 NP-DK50 DE-22_05-2023
CONTENTS
1. GENERAL INFORMATION ...................................................................................................... 5
INTENDED USE .................................................................................................................. 5
CONTRAINDICATIONS AND SIDE-EFFECTS ................................................................... 5
OPERATOR'S RESPONSIBILITY FOR PATIENT SAFETY ................................................ 5
MARKINGS ......................................................................................................................... 5
WARNINGS ........................................................................................................................ 5
General safety warnings ............................................................................................ 6
Electrical system safety warnings .............................................................................. 6
WARNING NOTICES AND SYMBOLS ................................................................................ 7
USE ..................................................................................................................................... 8
STORAGE AND TRANSPORT ........................................................................................... 8
2. EQUIPMENT DESCRIPTION ................................................................................................... 9
3. TECHNICAL DATA .................................................................................................................11
4. OPERATION ...........................................................................................................................13
INSTALLATION AND FIRST OPERATION ........................................................................13
Removal of transport stabilizers ...............................................................................13
Wheel installation .....................................................................................................14
Compressed air connection ......................................................................................14
Electrical connection.................................................................................................14
First operation ..........................................................................................................15
Accessories ..............................................................................................................16
OPERATION ......................................................................................................................16
Switching the compressor on ....................................................................................16
Running the compressor ..........................................................................................17
Cleaning and replacing filters ...................................................................................17
Cleaning the compressor ..........................................................................................17
5. MAINTENANCE ......................................................................................................................18
REPAIRS AND SERVICE...................................................................................................18
Cover removal ..........................................................................................................18
MAINTENANCE SCHEDULE .............................................................................................19
Safety valve check ...................................................................................................19
Checking tightness of joints and inspecting the equipment .......................................20
Replacing filter elements ..........................................................................................20
Setting output air pressure .......................................................................................21
Cleaning the pressure regulator ...............................................................................21
Replacing the suction filter (18) ................................................................................21
Stabilizing the compressor before shipping ..............................................................22
SHUTDOWN ......................................................................................................................22
EQUIPMENT DISPOSAL ...................................................................................................22
6. TROUBLESHOOTING.............................................................................................................23
7. SPARE PARTS .......................................................................................................................24
8. ELECTRIC AND PNEUMATIC DIAGRAMS ............................................................................25
WIRING DIAGRAM ............................................................................................................25
PNEUMATIC DIAGRAM .....................................................................................................26
9. ELECTROMAGNETIC COMPATIBILITY DECARATION ........................................................27
10. ANNEX .................................................................................................................................. 115
INSTALLATION RECORD................................................................................................ 115
GENERAL INFORMATION
NP-DK50 DE-22_05-2023 5
1. GENERAL INFORMATION
INTENDED USE
The EKOM DK50 DE is a medical air compressor that supplies clean, oil-free
compressed air for use with medical ventilators.
CONTRAINDICATIONS AND SIDE-EFFECTS
There are no contraindications or side-effects known.
OPERATOR'S RESPONSIBILITY FOR PATIENT SAFETY
The Installation, Operation and Maintenance Manual is an integral part of the
equipment and must be kept with the compressor. Careful review of this
manual will provide information necessary for correct operation of the
equipment.
Rx only
US Federal law restricts the sale of this device by or on the order of a
physician.
MARKINGS
Products marked with the CE mark of compliance meet the safety
requirements of the European Union (93/42/EEC).
WARNINGS
The safety of operating personnel and trouble-free operation of the
equipment are ensured only if original parts are used. Only accessories
and spare parts mentioned in the technical documentation or expressly
approved by the manufacturer can be used.
If any other accessories or consumable materials are used, the
manufacturer cannot be held responsible for the safe operation and
functionality of the equipment.
The warranty does not cover damages resulting from the use of
accessories or consumable materials other than those recommended by
the manufacturer.
The manufacturer assumes responsibility for the safety, reliability and
function of the equipment only if:
- Installation, calibration, amendments, extensions and repairs are
performed by the manufacturer, one of its representatives or a service
provider authorized by the manufacturer
- The equipment is used in accordance with the Installation, Operation and
Maintenance Manual
GENERAL INFORMATION
6 NP-DK50 DE-22_05-2023
The Installation, Operation and Maintenance Manual accurately describes
the design of the compressor and its compliance with safety and technical
standards. The manufacturer reserves all rights to its wiring diagrams,
procedures and names.
This user manual is the original instructions. Translation is performed in
accordance with the best available knowledge.
General safety warnings
The equipment is designed to operate safely when used correctly. Please
note the following safety measures to avoid injury or damage.
Equipment operation must comply with all local codes and regulations.
Original packaging should be kept for the possible return of the unit. Only
original packaging ensures optimal protection of the equipment during
transport. If it is necessary to return the equipment during the warranty
period, the manufacturer is not liable for damages caused by incorrect
packaging.
The user must immediately notify the supplier if any problem occurs during
the use of the equipment.
This product is not intended for use in areas where there is a risk of an
explosion. Do not operate the compressor in the presence of flammable
anesthetics.
Never feed oxygen or nitrous oxide into the compressor. Compressor
components are not approved for oxygen or nitrous oxide use.
The compressor must not be used for supplying air to the medical air
central distribution pipeline system.
Electrical system safety warnings
The equipment must be connected to ground. I order to assure proper
grounding, connect the compressor to a receptacle marked "hospital
grade."
Before the compressor is plugged in, make sure that the voltage and
frequency of the mains specified on the equipment are the same as the
power mains.
Before operating, check for possible damage to the equipment and any
connections. Damaged pneumatic and electrical lines must be replaced
immediately.
If a technical failure occurs, immediately disconnect the equipment from
the mains (pull out the main power plug).
During repairs and maintenance, ensure that:
- The main power plug is removed from the power socket
- Compressed air lines are disconnected
- All pressure has been released from the air tank
Only a qualified technician can install this equipment.
GENERAL INFORMATION
NP-DK50 DE-22_05-2023 7
WARNING NOTICES AND SYMBOLS
The following symbols are used for important information in the Installation,
Operation and Maintenance Manual and on packaging and the product:
Attention, see instructions for use
Caution, risk of electric shock
Refer to instruction manual
CE mark of compliance
Medical device
Serial number
Article number
Unique Device Identifier
Compressor is controlled automatically and may start without
warning
Caution, hot surface
Handling mark on package FRAGILE
Handling mark on package THIS SIDE UP
Handling mark on package KEEP DRY
Handling mark on package TEMPERATURE LIMITATIONS
Handling mark on package LIMITED STACKING
Mark on package RECYCLABLE MATERIAL
Ground connection
Equipotentiality
GENERAL INFORMATION
8 NP-DK50 DE-22_05-2023
Fuse
Condensate drain
Never dispose of the battery with common household waste.
USE
The equipment can be installed and operated only in a dry, ventilated and
dust-free area. Climatic conditions for operation - see Technical data.
The compressor must stand on a flat and stable base.
The compressor must not be exposed to rain. The equipment must not be
used in humid or wet environments. Never use the compressor in the
presence of flammable liquids or gases.
Before connecting the compressor to respiration equipment, make sure
that it meets the requirements of the respiration equipment. Refer to the
Technical data for this purpose.
Any use other than the compressor‘s intended use is not considered to be
safe. The manufacturer is not responsible for any damages that result if
the compressor is used for any other purpose. Risk is exclusively
assumed by the operator/user.
STORAGE AND TRANSPORT
The compressor is shipped from the factory in transport packaging with the
pump stabilized, protecting it from damage during transport.
For transport, always use the original packaging and secure the compressor
in the upright position.
Protect the compressor from humidity, contamination and extreme
temperatures during transport and storage. A compressor in its original
packaging should be stored in a warm, dry and dust-free area.
Keep the packaging material, if possible. If not, dispose of the packaging
material in an environmentally-friendly way. Cardboard can be recycled.
Before moving or transporting the compressor, the pressure in the air tank
and hoses must be released and any condensed water must be drained.
Secure the motor to prevent movement before shipping.
Prior to transport it is necessary to secure the motor inside the
compressor (Chapter 5).
EQUIPMENT DESCRIPTION
NP-DK50 DE-22_05-2023 9
2. EQUIPMENT DESCRIPTION
1. Switch, main connector, fuses
2. Hour counter
3. OUT compressed air output
4. Outlet pressure gauge (accuracy ± 5%)
5. WALL input of compressed air from central distribution (Auxiliary
equipment)
6. Suction filter
7. Equipotential (ground) pin
8. Condensate tank
9. Power indicator
10. Socket for the electrical cord
11. Compressor
12. Safety valve
13. Air tank
14. Filter (40 μm) and water trap
15. Pressure regulator
16. Filter (5 μm) and water trap
17. Control electronics
18. Suction filter
19. Cooler
The compressor contains an oil-free piston (11) driven by a low-maintenance single-phase
electric motor. Compressed air is cooled in the cooler (19) where condensed water is
separated into a separate tank (8). Incoming air passes through two filters (6,18)
undergoing double filtration as it passes through the system (14,16). Constant pressure at
the output is maintained by the pressure regulator (15). The built-in air tank (13) enables
peak air consumption of 200 L/min.
Backup ventilator compressed air source (applicable to compressors with WALL
connection)
Compressor air inlet (WALL) (5) is connected to terminal unit of the hospital central air
distribution through built-in non-return valve. Medical ventilator is connected to the
compressor outlet air connector. The compressor remains in STANDBY mode when the
central distribution air pressure is higher than compressor outlet air pressure. The air flows
from central distribution line through compressor air inlet (WALL), non-return valve and
compressor air outlet to the ventilator.
If the air pressure in central distribution drops below compressor outlet air pressure, the
controller automatically starts up the compressor, which delivers air directly to ventilator.
Air delivery from the compressor to the central air distribution pipeline is prevented by non-
return valve.
Primary ventilator compressed air source
The compressor continuously supplies compressed air directly to the medical ventilator
when the central compressed air distribution line is not connected to compressor air inlet
(WALL).
If air consumption is zero, the device switches to STANDBY mode.
The compressor must not be used for supplying air to the medical air
central distribution pipeline system. Do not connect compressor air
outlet to the central distribution pipeline system.
EQUIPMENT DESCRIPTION
10 NP-DK50 DE-22_05-2023
TECHNICAL DATA
NP-DK50 DE-22_05-2023 11
3. TECHNICAL DATA
DK50 DE
basic
standard
advanced
L/min
40 / 32*
50 / 40*
60 / 50*
Peak flow
200** L/min ( 7 Cft/min )
V/Hz/A
230/50 / 2.8
230/60 / 2.8
120/60 / 5.6
100/50-60 / 5.6
230/50 / 2.8
230/60 / 2.8
120/60 / 5.6
100/50-60 / 5.6
230/50 / 3.3
230/60 / 3.9
120/60 / 6
100/50-60 / 6
m
5
Pressure dew point at 40 L/min
( 1.4 Cft/min ), 3 bar( 43.5 psig ), 20°C (68°F)
5°C (9°F) below the ambient temperature
DISS 1160-A (3/4"-16 UNF)
dB(A)
@50Hz
49
50
51
Mode of operation
Continuous - S1
Separation of condensed water
Automatic
Output pressure
3.0 bar (43 psig)
Adjustable to max. 3.5 bar (51 psig)
Automatic start up pressure (backup***)
Decrease in pressure in the central distribution line under
2.8 bar ( 40.6 psig)
2 L (0.61gall UK)
Pressure range
5 bar ( 72.5 psig ) 6.5 bar ( 94 psig)
Operating pressure of safety valve
7 bar ( 101.5 psig )
Adjustment of output air pressure
Pressure regulator
w x d x h
440 x 350 x 420 mm (17.5 x 14x 16.5 in )
w x d x h
470 x 380 x 505 mm (18.5 x 15x 20 in )
w x d x h
510 x 480 x 470 mm ( 20 x 19 x 18.5 in )
Net weight
34 kg ( 75 lbs )
Net weight of compressor witsh wheels
36 kg ( 79 lbs )
Gross weight
42 kg ( 93 lbs )
Gross weight of compressor with wheels
43 kg ( 95 lbs )
Implementation according to EN 60601-1, EN
12021
Class I.
Classification acc. to MDD 93/42 EEC, 2007/47
EC
II b
* ) With membrane dryer (optional)
**) With 0.6 bar drop (8.7 psig)
***) Extra equipment
TECHNICAL DATA
12 NP-DK50 DE-22_05-2023
Climatic conditions for storage and transport
Temperature –25°C to +55°C (-13°F to +131°F), 24 hrs +70°C (+158°F)
Relative air humidity 0% to 100% (with condensation)
Climatic conditions for operation
Temperature +15°C to +40°C (+59°F to +104°F)
Relative air humidity up to +95%
IPX0 Rating
FAD efficiency correction for differences in elevation
FAD correction table
Elevation [mamsl]
0 - 1500
1501 - 2500
2501 - 3500
3501 - 4500
FAD [l/min]
FAD x 1
FAD x 0.8
FAD x 0.71
FAD x 0.60
FAD efficiency refers to conditions at an elevation of 0 mamsl:
Temperature: 20°C
Atmospheric pressure: 101325 Pa
Relative humidity: 0%
OPERATION
NP-DK50 DE-22_05-2023 13
4. OPERATION
INSTALLATION AND FIRST OPERATION
Do not use the compressor immediately after unpacking as it will not
adjust to the ambient temperature.
Only qualified personnel can install the compressor and put it into
operation for the first time. The installer shall train the operating
personnel in the use and routine maintenance of the equipment.
Installation and training of personnel should be acknowledged by the
installer‘s signature on the installation certificate.
Prior to start-up, the four screws for transport stabilization must be
removed. If the compressor is switched on without removing the
stabilizers, the compressor could be permanently damaged.
The compressor does not contain a backup power supply.
Never block the suction filter on the backside or the vent outlets on the
top of the equipment.
If the compressor is equipped with a main source of air, the standby air
source must be available.
ANY MODIFICATION OF THIS EQUIPMENT IS FORBIDDEN!
If this equipment is used nearby other instruments, the equipment must
be observed in order to verify normal operations in the configuration it
will be used.
Instruments may be affected electro-magnetically!
You may notice a “new product” odour when you first place the
product into service (for a short period of time). This odour is
temporary and does not impede the normal use of the product. Ensure
the space is properly ventilated after installation.
Removal of transport stabilizers
On the bottom of the unit are four M6
stabilizing screws that must be removed
before start-up. They are marked with a
red warning washer.
Please retain the stabilizing screws for
future transport of the compressor.
OPERATION
14 NP-DK50 DE-22_05-2023
Wheel installation
Wheels must be installed following the figure below for versions of the
compressor with wheels.
Compressed air connection
The given appliance or respiratory apparatus should be connected to the
OUT (3) quick coupler, as this is the outlet for compressed air.
Applicable to compressors with WALL connection (5):
Compressed air from the central distribution line is connected via the WALL
(5) quick coupling as this is the inlet for compressed air, if applicable. Air
from the distribution system is automatically connected through the
compressor to the OUT outlet port. In this configuration, the compressor
serves as a backup source of compressed air. If the air pressure from the
central distribution system is reduced, the compressor automatically switches
on and there is no interruption in the supply of continuously pressurized air at
the outlet of the compressor.
Air delivery from the compressor to the central air distribution pipeline is
prevented by non-return valve.
Please note that the air supply from the central distribution line
entering the compressor must be medical grade air (particulate size,
humidity.) The compressor does not modify the air from the central
distribution line.
The hose connecting the compressor to the respiration equipment
must not pass through a cold environment i.e. placed on the ground. It
should be as short as possible with no kinks (this may cause water to
condense inside the hose).
The compressor must not be used for supplying air to the medical air
central distribution pipeline system.
Electrical connection
The compressor comes with a plug containing an appropriate
protective contact (ground.) Adhere to local electrical regulations. The
voltage and frequency of the mains must comply with the
specifications on the data label.
OPERATION
NP-DK50 DE-22_05-2023 15
The electrical cord must not be stressed or have any tension exerted
upon it, and must always be free.
The socket must be accessible for safety reasons so that the equipment
can be safely disconnected from the power supply in case of an
emergency.
The relevant current circuit must be protected.
Connection of the ground connection (7) to other equipment must adhere
to local electrical regulations.
Fasten the electrical cord through the holder (10).
First operation
Make sure that the stabilizing screws used during transport were
removed.
Check that the connection to the compressed air supply is correct.
Check for proper connection to the main power supply.
Switch on the pressure switch (1) to position “I”.
The compressor will be operational once it is put into operation. If there is no
air consumption at the output, the device switches to STANDBY mode. When
air consumption resumes, the compressor automatically switches on.
During operation, the device drains off the entrapped condensed water from
pneumatic circuits via automatic filter separators into a tank.
The following applies to device with additional equipment WALL (5)
coupling:
After the compressor is put into operation, it shall work in one of the following
modes depending upon the pressure level in the central distribution and upon
air consumption.
STANDBY When there is sufficient pressure in the central air distribution
line, the main switch is on, the compressor is idle. The device operates as a
standby source of air as it checks the pressure in the central air distribution
line and, if the pressure falls, the compressor switches on.
When pressure is low in the central air distribution line, or if the device is
not connected to the central distribution line, the compressor is running. If
there is no air consumption at the outlet, the device switches to STANDBY
mode. When air consumption resumes, the compressor automatically
switches on.
During operation, the device drains the trapped condensed water from
pneumatic circuits via automatic filter separators into a separate tank.
OPERATION
16 NP-DK50 DE-22_05-2023
Accessories
Trolley SD 30 (24)
The trolley accommodates the compressor, a fan and a humidifier. Large
casters allow for mobility and braking, guided by means of an ergonomic
handle. A wide base ensures the stability of the entire assembly.
Specifications
Supply No.: 602021222-000
Dimensions: 535x575x1070
Maximal load-carrying capacity with load
in the upper plate axis (1):
Trolley without compressor - 25kg
Trolley with compressor - 30kg
The manufacturer is not liable for any damage
resulting from excessive loads on the equipment!
The supplier is obliged to ensure the acceptability of a load delivered
by a supplied accessory.
It is forbidden to lean or press against an installed accessory for
support!
The equipment must be lifted when travelling over an obstacle!
Supporting equipment must be disassemble before transport!
The maximum load on the upper trolley plate is 30 kg!
Humidifier holder (25)
This clamp is used to attach the Fischer & Paykel humidifier to the trolley in
its proper position.
Supply No. 604031175-000
OPERATION
The equipment may only be operated by the trained staff!
In case of emergency, switch the equipment off at the switch and pull
out the main power plug.
Switching the compressor on
The compressor is switched on at the main power switch (1) by putting it in
position "I".
OPERATION
NP-DK50 DE-22_05-2023 17
Running the compressor
Compressor operation is automatic. The compressor runs permanently,
when air consumption is lower than the set point; once set point is reached
the operation is switched to idle run. The pressure gauge displays the value
of output pressure. When air consumption stops, the compressor shall switch
off after around 30 seconds and switch on after air consumption is restored.
The condensed liquid drains into a tank (8) at the back of the device. When
the tank fills up, it must be emptied.
Cleaning and replacing filters
At least once a week take out and clean the suction filter (6) located on the
back side. Wash the filter in warm soapy water, rinse thoroughly and allow it
to dry. Insert the clean filters so that the intake openings are completely
covered by the filters. A replacement suction filter (6) No. 025000018 is
included in the basic equipment for the product.
Cleaning the compressor
To clean the compressor, use a detergent that contains no abrasives,
chemical solvents or other corrosive agents.
MAINTENANCE
18 NP-DK50 DE-22_05-2023
5. MAINTENANCE
REPAIRS AND SERVICE
Warranty and extended warranty repairs are to be completed by the
manufacturer or a service provider authorized by the manufacturer.
The manufacturer reserves the right to modify the equipment in any
way that will not alter the function or the operation of the equipment.
Only a qualified technician or the Customer Service Department of the
manufacturer may perform repairs that go beyond routine maintenance.
Use only spare parts and accessories approved by the manufacturer.
Prior to any maintenance or repairs, switch off the compressor and
disconnect it from the mains (pull out the main power plug).
Air pump components (head, cylinder, pressure hose, etc.) are very hot
during and shortly after compressor operation do not touch these
components!
Allow the equipment to cool down before maintenance, service or
connecting/disconnecting the compressed air supply!
Automatic start: when pressure in the pressure tank decreases below
the switch-on pressure, the compressor automatically switches on. The
compressor automatically switches off when pressure in the air tank
reaches the switch-off pressure.
Cover removal
- Unscrew the 6 screws from the rear cover
- Disconnect the grounding wire
- Remove the rear cover
- Unscrew the 4 screws from the rear of the main cover
- Disconnect the grounding wire
- Remove the main cover
- Reassemble using the opposite order
MAINTENANCE
NP-DK50 DE-22_05-2023 19
MAINTENANCE SCHEDULE
Notice!
The operating entity is obliged to ensure that all tests of the equipment are carried
out repeatedly at least once within every 24 months (EN 62353) or in intervals as
specified by the applicable national legal regulations. A report must be prepared on
the results of the tests (e.g.: according to EN 62353, Annex G), including the
measurement methods used.
Maintenance
Page
Time interval
To be
performed by
Clean suction filter
17
At least once a week
Staff
Safety valve check
19
Once per year
Qualified expert
Replace filters in filter
20
Every 4000 hours
Qualified expert
Test the tightness of joints and
inspect the equipment
20
After two years
Qualified expert
Replace piston complete with piston
rings, O-rings and bearing
Service
documentation
Every 8000 hours
Qualified expert
Replace suction filter
21
Every 8000 hours
Qualified expert
Perform “Repeated Test” according
to EN 62353
19
1 x 2 years
Qualified expert
Safety valve check
The safety valve is adjusted to 7 bar by the manufacturer, then tested
and marked. It must not be readjusted.
Only allow the safety valve freely
puff for a short period of time.
Never use the safety valve to release pressure in the air tank. This could
damage the safety valve.
MAINTENANCE
20 NP-DK50 DE-22_05-2023
Checking tightness of joints and inspecting the equipment
Test for leakage:
Disconnect the outlet hose from the quick coupling OUT (3); compressed
air consumption will stop.
Pressurize the air tank by releasing some air via a connected device.
With the compressor off, wait at least 10 minutes before checking the
pressure level.
If the pressure has dropped, use soapy water to find the leaky joint. Start
with the joints outside the case, and then test the ones inside the case
after removing the covers.
Tighten or re-seal joints as necessary.
Inspect the equipment:
Check the status of the compressor motor for balanced operation or noise.
Check the condition of the hangers above the pump.
Check fan functionality.
Check the condition of filters (14,16). With regular operation, condensate
drains automatically. Verify this function by comparing the level of
condensate in the tank to the scale line indicating the maximum level. If
the level of condensate is above the scale line, replace the faulty parts.
Examine the pump for:
- Defects in the crankcase
- Free movement of the crank shaft
If necessary, replace any faulty parts.
Replacing filter elements
Before beginning, evacuate all compressed air from the tank, reducing
the tank pressure to zero and disconnect the equipment from the power
mains.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120

EKOM DK50 DE Používateľská príručka

Typ
Používateľská príručka

v iných jazykoch