Hach pHD Sensor Používateľská príručka

Typ
Používateľská príručka
DOC023.98.80089
pHD Sensor
07/2016, Edition 3
User Manual
Bedienungsanleitung
Manuale utente
Manuel d'utilisation
Manual del usuario
Manual do utilizador
Návod k použití
Brugervejledning
Gebruikershandleiding
Instrukcja obsługi
Bruksanvisning
Ръководство за потребителя
Felhasználói kézikönyv
Manual de utilizare
Руководство пользователя
Kullanıcı Kılavuzu
Návod na obsluhu
Navodila za uporabo
User Manual
Εγχειρίδιο χρήσης
 
English..............................................................................................................................3
Deutsch.......................................................................................................................... 20
Italiano............................................................................................................................ 39
Français......................................................................................................................... 57
Español.......................................................................................................................... 75
Português...................................................................................................................... 93
Čeština......................................................................................................................... 111
Dansk............................................................................................................................129
Nederlands................................................................................................................. 147
Polski............................................................................................................................ 165
Svenska....................................................................................................................... 183
български................................................................................................................... 201
Magyar......................................................................................................................... 220
Română....................................................................................................................... 238
Русский........................................................................................................................256
Türkçe...........................................................................................................................275
Slovenský jazyk......................................................................................................... 293
Slovenski..................................................................................................................... 311
Hrvatski........................................................................................................................ 329
Ελληνικά...................................................................................................................... 347
.............................................................................................................................. 366
2
Specifications
Specifications are subject to change without notice.
Specification Details
Measurement range (pH) 2.5 to 12.5 pH
Measurement range (temperature) -5 to 95 °C (23 to 203 °F)
Resolution 0.01 or 0.1 pH
Temperature compensation 300 ohm NTC thermistor
Stability (analyzer only) 0.03 pH per 24 hours, non-cumulative
Sensitivity Less than 0.005 pH
Maximum probe immersion depth/
pressure
6.9 bar at 105 °C (100 psi at 221 °F)
Maximum flow rate 3 m (10 ft) per second
Power requirements 5 VDC, 1 mA (supplied by controller)
Operating temperature -5 to 95 °C (23 to 203 °F)
Cable lengths/type 6 m (20 ft), 5 conductor (plus two isolated shields) cable with XLPE
(cross-linked polyethylene) jacket; rated to 150 °C (302 °F)
Maximum transmission distance 914 m (3000 ft)
Calibration methods Initial 2-point calibration using 2 buffers and then option to use 1-point or
2-point (slope) calibration using samples or buffers
Interfaces Modbus from gateway
Material Ryton
®
(PVDF) body, salt bridge of matching material with Kynar
®
junction, glass process electrode, titanium ground electrode, and Viton
®
O-ring seals
General Information
In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer’s website.
Safety information
Please read this entire manual before unpacking, setting up or operating this equipment. Pay
attention to all danger and caution statements. Failure to do so could result in serious injury to the
operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired, do not use or install this
equipment in any manner other than that specified in this manual.
Use of hazard information
D A N G E R
Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.
W A R N I N G
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious
injury.
English 3
C A U T I O N
Indicates a potentially hazardous situation that may result in minor or moderate injury.
N O T I C E
Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special
emphasis.
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument
could occur if not observed. A symbol on the instrument is referenced in the manual with a
precautionary statement.
This is the safety alert symbol. Obey all safety messages that follow this symbol to avoid potential
injury. If on the instrument, refer to the instruction manual for operation or safety information.
This symbol indicates that a risk of electrical shock and/or electrocution exists.
This symbol indicates the presence of devices sensitive to Electro-static Discharge (ESD) and
indicates that care must be taken to prevent damage with the equipment.
Electrical equipment marked with this symbol may not be disposed of in European public disposal
systems after 12 August of 2005. In conformity with European local and national regulations (EU
Directive 2002/96/EC), European electrical equipment users must now return old or end-of-life
equipment to the Producer for disposal at no charge to the user.
Product overview
This sensor is designed to work with the digital gateway for the CLF10sc and CLT10sc Reagentless
Chlorine Analyzer and one of the sc series controllers for data collection and operation.
This sensor has an internal temperature sensor (thermistor). The temperature measurement signal is
used internally by the sensor for automatic temperature compensation and is shown on the
controller.
Theory of operation
pH is the negative logarithm of the hydrogen ion activity and a measure of the acidity or alkalinity of a
solution.
pH is normally measured with a glass electrode and a reference electrode. The glass electrode acts
as a transducer which converts chemical energy (the hydrogen ion activity) into an electrical energy
(measured in millivolts). The reaction is balanced and the electrical circuit is completed by the flow of
ions from the reference solution to the solution under test.
The electrode and reference solution together develop a voltage (emf) whose magnitude depends on
the type of reference electrode, the internal construction of the glass electrode, the pH of the solution
and the temperature of the solution.
Product components
Refer to Figure 1 to make sure that all components have been received. If any of these items are
missing or damaged, contact the manufacturer or a sales representative immediately.
4
English
Figure 1 Sensor components
1 pHD sensor 2 Sealing hub for pH flow cell
Installation
C A U T I O N
Personal injury hazard. Only qualified personnel should conduct the tasks described in this section of the manual.
Install the sensor
The pH sensor must be installed in the flow cell, connected to the gateway and calibrated before use.
The sensor does not need to be conditioned. To install the sensor, refer to the illustrated steps.
English
5
1 2
3 4
6 English
Connect the sensor to the gateway
D A N G E R
Electrocution Hazard. High voltage wiring for the controller is conducted behind the high voltage barrier in the
controller enclosure. The barrier must remain in place except when installing modules, or when a qualified
installation technician is wiring for power, relays or analog and network cards.
W A R N I N G
Potential Electrocution Hazard. Always disconnect power to the instrument when making electrical
connections.
N O T I C E
Potential Instrument Damage. Delicate internal electronic components can be damaged by static
electricity, resulting in degraded performance or eventual failure.
Pre-requisites: Make sure the sensor cable is routed through the sealing hub and then the lock ring
for the pH flow cell before doing this procedure. Refer to Install the sensor on page 5 for the
illustrated steps.
To connect the sensor to the gateway, refer to the illustrated steps and Table 1.
Table 1 pHD sensor wire connections
Connector Pin Signal Sensor wire
J4 WHT -5 VDC White
SHLD Solution ground Clear (2 wires)
BLK Temp – Black
YEL Temp + Yellow
GRN (TB1) 1 Reference Green
2 Reference
RED (TB2) 1 Active/Measuring
2 Active/Measuring Red
English 7
1
2 3
8 English
4
5 6
English 9
Operation
Guidelines for operation
C A U T I O N
Personal injury hazard. If the pH process electrode breaks, handle the sensor very carefully to prevent injury.
Before the pH sensor is placed in operation, remove the protective cap to expose the process
electrode and salt bridge. Save the protective cap for future use.
The process electrode at the pH sensor tip has a glass bulb, which can break. Do not subject this
electrode to abrupt impact or other mechanical abuse.
For short-term storage (when the sensor is out of the process for more than one hour), fill the
protective cap with pH 4 buffer or distilled water and put the cap back on the sensor. Keep the
process electrode and salt bridge moist to avoid slow response when the sensor is returned to
operation.
For extended storage, repeat the short-term storage procedure every 2 to 4 weeks, depending on
the environmental conditions.
User navigation
Refer to the controller documentation for keypad description and navigation information.
Configure the sensor
Use the Configure menu to enter identification information and display options for the sensor and to
change options for data handling and storage.
1. Push the MENU key and select Sensor Setup, Configure.
Option Description
EDIT NAME Changes the name that corresponds to the sensor on the top of the measure screen.
The name is limited to 10 characters in any combination of letters, numbers, spaces or
punctuation. The default name is the serial number of the sensor.
SELECT PARAM. Customizes the options for sensor data handling and storage. Refer to Select
temperature parameters on page 10 and Select pH parameters on page 11.
RESET DEFAULTS Sets the configuration menu to the default settings. All sensor information is lost.
Select temperature parameters
1. Select the type of chlorine sensor used - Total CL2 or Free CL2.
2. Select Yes.
3. Select DIFF PH.
4. Select Temperature.
5. Customize the options:
Option Description
SELECT UNITS Sets the units for the temperature measurements-°C (default) or °F.
FILTER Sets a time constant to increase signal stability. The time constant calculates the average
value during a specified time-0 (no effect, default) to 60 seconds (average of signal value
for 60 seconds). The filter increases the time for the sensor signal to respond to actual
changes in the process.
LOG SETUP Sets the time interval for data storage in the data log-10, 30 seconds, 1, 5, 15 (default),
60 minutes.
10 English
Select pH parameters
1. Select the type of chlorine sensor used - Total CL2 or Free CL2.
2. Select Yes.
3. Select DIFF PH.
4. Select pH.
5. Customize the options:
Option Description
DISPLAY
FORMAT
Sets the number of decimal places that are shown on the measure screen-XX.XX or
XX.X
FILTER Sets a time constant to increase signal stability. The time constant calculates the
average value during a specified time-0 (no effect, default) to 60 seconds (average of
signal value for 60 seconds). The filter increases the time for the sensor signal to
respond to actual changes in the process.
LOG SETUP Sets the time interval for data logging-10, 30 seconds, 1,5, 15 (default), 60 minutes.
Calibrate the sensor
About sensor calibration
The sensor characteristics slowly shift over time and cause the sensor to lose accuracy. The sensor
must be calibrated regularly to maintain accuracy. The calibration frequency varies with the
application and is best determined by experience.
Recalibrate the sensor whenever disconnected from power and removed from water.
Temperature calibration procedure
One measurement is required for temperature calibration of this sensor. The measurement is made
with the pH sensor in a beaker that contains a sample or reference solution or with the pH sensor
installed in the flow cell.
1. To calibrate the temperature with the pH sensor in a beaker:
a. Put the sensor in the sample or reference solution.
b. Make sure that the sensor is at least half immersed in the liquid (Figure 2 on page 12).
c. Stir the sensor to remove bubbles.
d. Wait for the sensor and solution temperature to equalize. This can take 30 minutes or more.
2. To calibrate the temperature with the pH sensor in the flow cell, install the pH sensor in the flow
cell and turn on the flow. Wait at least 30 minutes after flow is started for the temperature
readings of the pH sensor to stabilize.
3. Push the MENU key and select Sensor Setup, Calibrate, Temperature, Temp Cal.
4. If the passcode is enabled in the security menu for the controller, enter the passcode.
The controller shows "Stabilizing" until the temperature measurement stabilizes and then shows a
temperature measurement.
5. Select the option for the output signal during calibration:
Option Description
ACTIVE The instrument sends the current measured output value during the calibration procedure.
HOLD The sensor output value is held at the current measured value during the calibration procedure.
TRANSFER A preset output value is sent during calibration. Refer to the controller user manual to change
the preset value.
6. Measure the temperature of the sample or reference solution with a secondary verification
instrument (such as an NIST traceable thermometer).
7. Use the arrow keys to enter the measured value and push ENTER.
English
11
8. Review the calibration result:
Passed—the sensor is calibrated and ready to measure samples. The offset value is shown.
Failed—the calibration offset is outside of accepted limits. Refer to Troubleshooting
on page 16 for more information.
9. If the calibration passed, push ENTER to continue.
10. If the option for operator ID is set to Yes in the Calibration Options menu, enter an operator ID.
Refer to Change the calibration options on page 14.
11. On the New Sensor screen, select whether the sensor is new:
Option Description
YES The sensor was not calibrated previously with this instrument. The days of operation and previous
calibration curves for the sensor are reset.
NO The sensor was calibrated previously with this instrument.
12. Return the sensor to the process and push ENTER.
The output signal returns to the active state and the measured sample value is shown on the
measure screen.
Note: If the output mode is set to hold or transfer, select the delay time when the outputs return to the active
state.
pH calibration procedure
Pre-requisites: Do a temperature calibration before doing a pH calibration. The accuracy of the pH
measurement depends on the accuracy of the temperature measurement.
One or two measurements are required for pH calibration of this sensor. Measurements are made
with the pH sensor in a beaker that contains a sample or reference solution or with the pH sensor
installed in the flow cell.
Note: The pH sensor should first be calibrated with a reference solution(s) in a beaker. Then the pH sensor can be
calibrated with a sample(s) in a beaker or in the flow cell.
The pH can be calibrated with 1 or 2 reference solutions or samples (1-point or 2-point calibration).
Calibration adjusts the sensor reading to match the value of a reference solution(s) or sample(s).
A calibration is done by putting the pH sensor in a reference solution or sample with a known pH
value and then entering that known value into the controller. A buffer calibration identifies the buffer
table corresponding to the chosen buffer and automatically calibrates the probe after it stabilizes.
1. To calibrate the pH sensor in a beaker:
a. Put the sensor in the reference solution or sample.
b. Make sure that the sensor is at least half immersed in the liquid (Figure 2).
c. Stir the sensor to remove bubbles.
d. Wait for the sensor and solution temperature to equalize. This can take up to 30 minutes.
Figure 2 Sensor in reference solution or sample
2. To calibrate the pH sensor in the flow cell, install the pH sensor in the flow cell and turn on the
flow.
12
English
3. Push the MENU key and select Sensor Setup, Calibrate, pH.
4. Select the type of calibration:
Option Description
2 POINT
BUFFER
Use 2 buffers for calibration, for example pH 7 and pH 4 (recommended method). The
buffers must be from the buffer set that is specified in the Cal Options menu (refer to
Change the calibration options on page 14).
1 POINT
BUFFER
Use 1 buffer for calibration, for example pH 7. The buffer must be from the buffer set that
is specified in the Cal Options menu (refer to Change the calibration options
on page 14).
2 POINT
SAMPLE
Use 2 samples of known pH value for calibration. Determine the pH value of samples with
a different instrument.
1 POINT
SAMPLE
Use 1 sample of known pH value for calibration. Determine the pH value of sample with a
different instrument.
5. If the passcode is enabled in the security menu for the controller, enter the passcode.
6. Select the option for the output signal during calibration:
Option Description
ACTIVE The instrument sends the current measured output value during the calibration procedure.
HOLD The sensor output value is held at the current measured value during the calibration procedure.
TRANSFER A preset output value is sent during calibration. Refer to the controller user manual to change
the preset value.
7. With the sensor in the first reference solution or sample, push ENTER.
The measured pH and temperature value is shown.
8. Wait for the value to stabilize and push ENTER
1
.
9. If using a sample, measure the pH value with a secondary verification instrument. Use the arrow
keys to enter the measured value and push ENTER.
Note: If a pH buffer not listed in the Cal Options menu is used, refer to the buffer bottle to find the pH value that
corresponds to the temperature of the buffer.
10. For a 2-point calibration:
a. If using a reference solution, remove the sensor from the first solution and rinse with clean
water.
b. Put the sensor in the next reference solution or sample and push ENTER.
The measured pH and temperature value is shown.
c. Wait for the value to stabilize. Push ENTER
1
.
d. If the solution is a sample, measure the pH value with a secondary verification instrument.
Use the arrow keys to enter the measured value and push ENTER.
Note: If a pH buffer not listed in the Cal Options menu is used, refer to the buffer bottle to find the pH value
that corresponds to the temperature of the buffer.
11. Review the calibration result:
Passed—the sensor is calibrated and ready to measure samples. The slope and/or offset
values are shown.
Failed—the calibration slope or offset is outside of accepted limits. Repeat the calibration with
fresh reference or sample solution. Refer to Troubleshooting on page 16 for more
information.
12. If the calibration passed, push ENTER to continue.
1
If the option for Auto Stab (auto stabilization) is set to Yes in the Calibration Options menu, the
screen will advance to the next step automatically. Refer to Change the calibration options
on page 14.
English 13
13. If the option for operator ID is set to Yes in the Calibration Options menu, enter an operator ID.
Refer to Change the calibration options on page 14.
14. On the New Sensor screen, select whether the sensor is new:
Option Description
YES The sensor was not calibrated previously with this instrument. The days of operation and previous
calibration curves for the sensor are reset.
NO The sensor was calibrated previously with this instrument.
15. Return the sensor to the process and push ENTER.
The output signal returns to the active state and the measured sample value is shown on the
measure screen.
Note: If the output mode is set to hold or transfer, select the delay time when the outputs return to the active
state.
Reset calibration to defaults
To remove a bad calibration, replace the user calibration settings with the default calibration settings
using the Calibrate menu. Then recalibrate the sensor when needed.
1. Push the MENU key and select Sensor Setup, Calibrate, [Select Sensor], Reset Defaults.
2. If the passcode is enabled in the security menu for the controller, enter the passcode.
3. Select Yes and push Enter.
Change the calibration options
The user can select buffer solutions for pH calibrations, set a calibration reminder, enable auto
stabilization during calibrations or include an operator ID with calibration data from the Cal Options
menu.
1. Push the MENU key and select Sensor Setup, Calibrate, [Select Sensor], Cal Options.
2. Customize the options:
Option Description
SELECT BUFFER For pH only-changes the set of buffer solutions that are recognized for calibration to pH
4.00, 7.00, 10.00 (default set) or DIN 19267 (pH 1.09, 4.65, 6.79, 9.23, 12.75)
Note: Other buffers can be used if the 1 point sample or 2 point sample option is
selected during calibration.
AUTO STAB For pH only-enables the system to accept measurement signal values during calibrations
and advance to the next step of the calibration when the system determines the
measurement signal has stabilized-On or Off (default). Enter a stabilization range-0.01 to
0.1 pH unit.
CAL REMINDER Sets a reminder for the next calibration in days, months or years.
OP ID ON CAL Includes an operator ID with calibration data—Yes or No (default). The ID is entered
during the calibration.
Data log
The controller provides one data log for each sensor. The data log stores the measurement data at
selected intervals (user configurable). The data log can be read out in a CSV format. For instructions
on downloading the logs, please refer to the controller user manual.
Refer to Select temperature parameters on page 10 and Select pH parameters on page 11 for
information about setting time intervals for data storage in the data log.
Modbus registers
A list of Modbus registers is available for network communication. Refer to the CD for more
information.
14
English
Maintenance
C A U T I O N
Personal injury hazard. Only qualified personnel should conduct the tasks described in this section of the manual.
Maintenance schedule
Maintenance task Frequency
Clean and inspect the sensor 90 days
(The pH sensor may need to be cleaned more often
depending on water quality.)
Replace the standard cell solution and salt bridge 3 to 6 months
Replace the sensor 4-5 years
Clean the sensor
W A R N I N G
Chemical hazard. Always wear personal safety protection in accordance with the Material Safety Data Sheet for
the chemical that is used.
Examine the sensor periodically for debris and deposits. Clean the sensor when there is a buildup of
deposits or when performance has degraded.
Pre-requisites: Prepare a mild soap solution with a non-abrasive dishwashing detergent that does
not contain lanolin. Lanolin leaves a film on the electrode surface that can degrade the sensor
performance.
1. Turn off the flow.
2. Loosen the lock ring and remove the pH sensor from the flow cell.
3. Rinse the sensor with a stream of clean, warm water. If debris remains, carefully wipe the entire
measuring end of the sensor with a clean, soft cloth to remove the loose contaminate buildup.
Then rinse with clean water.
4. Soak the sensor for 2 to 3 minutes in the soap solution.
5. Use a soft bristle brush and scrub the entire measuring end of the sensor, thoroughly cleaning the
electrode and salt bridge surfaces.
6. If surface deposits remain, soak the measuring end of the sensor in dilute acid, such as muriatic
acid (or other dilute acid), for a maximum of 5 minutes.
Note: The acid should be as dilute as possible, not stronger than 3% HCL. Experience will determine which
acid to use and the appropriate dilution ratio. Some stubborn coatings may require a different cleaning agent.
Contact technical support.
7. Rinse the sensor with water and return to the soap solution for 2 to 3 minutes to neutralize any
remaining acid.
8. Rinse the sensor with clean water.
9. Calibrate the sensor in a beaker using a reference solution(s).
10. Install the pH sensor in the flow cell and tighten the lock ring.
English
15
Troubleshooting
Test the sensor
Pre-requisites: Two pH buffers (pH 7 and pH 4 or pH 10) and a multimeter.
Note: If a calibration fails, clean the sensor and replace the salt bridge and standard cell solution, and then repeat
the calibration. Only test the sensor if the problem is not corrected by maintenance.
1. Put the sensor in a pH 7 buffer solution and wait for the temperature of the sensor and buffer to
reach room temperature.
2. Disconnect the yellow and black sensor wires from the gateway.
3. Measure the resistance between the yellow and black wires to verify the operation of the
temperature element. The resistance should be between 250 and 350 ohms at approximately
25 ºC.
If the temperature element is good, reconnect the wires to the gateway.
4. Push the MENU key and select Sensor Setup, Diag/Test, Signals. The pH reading should be
between –50 and + 50 mV.
5. Rinse the sensor with water and put in a pH 4 or pH 10 buffer solution. Wait for the temperature
of the sensor and buffer to reach room temperature.
6. Compare the mV reading in the pH 4 or 10 buffer to the reading in the pH 7 buffer. The reading
should differ by approximately 160 mV.
If the difference is less than 160 mV, call technical support.
Diagnostic and test menu
The diagnostic and test menu shows current and historical information about the chlorine analyzer.
Refer to Table 2. To access the diagnostic and test menu, push the MENU key and select Sensor
Setup, Diag/Test.
Table 2 DIAG/TEST menu
Option Description
GATEWAY INFO Shows the firmware version, driver version, serial number and boot version for the
controller and the types of sensors connected to the controller.
CAL DAYS Shows the number of days since the sensor was last calibrated.
CAL HISTORY Shows a list of the times when the sensor was calibrated. Push ENTER to scroll
through the entries and view a summary of the calibration data.
RST CAL HISTORY Resets the sensor calibration history. Requires passcode.
SIGNALS Shows the sensor measurement signal value in mV.
SENSOR DAYS Shows the number of days the sensor has been in operation.
RST SENSORS Resets the sensor days and calibration days to default. Requires passcode.
CALIBRATION Shows the slope and offset values for chlorine and pH. Shows the offset value for
the temperature.
16 English
Error list
Errors may occur for various reasons. The sensor reading on the measurement screen flashes. All
outputs are held when specified in the controller menu. To show the sensor errors, push the MENU
key and select Sensor Diag, Error List. A list of possible errors is shown.
Table 3 Error list for the sensor
Error Description Resolution
CL CAL REQD A chlorine calibration and/or pH calibration is
required
The chlorine and/or pH measurement has
changed enough to cause a Cal Watch alarm
to occur. Refer to the chlorine sensor manual
for information on Cal Watch alarms.
Calibrate the chlorine sensor and/or pH
sensor.
PH TOO LOW The pH value is less than 0 pH Calibrate or replace the pH sensor.
PH TOO HIGH The pH value is more than 14 pH
PH SLOPE FAIL The slope is outside of the -45 to -65 mV/pH
range
Clean the pH sensor, then repeat the
calibration with a fresh buffer or sample, or
replace the sensor.
PH OFFSET FAIL The offset is outside of ±60 mV Clean the pH sensor and replace the salt
bridge and standard cell solution and then
repeat the calibration with a fresh buffer or
sample, or replace the sensor.
TEMP TOO LOW The temperature is less than 0 °C Calibrate the temperature or replace the pH
sensor.
TEMP TOO HIGH The temperature is more than 100 °C
TEMP FAIL The offset is higher than 5.0 °C or lower than
-5.0 °C
Calibrate the temperature or replace the pH
sensor.
Warning list
A warning does not affect the operation of menus, relays and outputs. A warning icon flashes and a
message is shown on the bottom of the measurement screen. To show the sensor warnings, push
the MENU key and select Sensor Diag, Warning List. A list of possible warnings is shown in Table 4.
Table 4 Warning list for the sensor
Warning Description Resolution
CL CAL RECD A chlorine and/or pH calibration is recommended
The chlorine and/or pH measurement has changed
enough to cause a Cal Watch warning alarm to occur.
Refer to the chlorine sensor manual for information on
Cal Watch alarms.
Calibrate the chlorine sensor
and/or pH sensor.
PH CAL RECD A pH calibration is recommended
pH calibration data is not available (sensor with default
calibration data)
Calibrate the pH sensor.
TEMP CAL RECD A temperature calibration is recommended
Temperature calibration data is not available (sensor
with default calibration data)
Calibrate the temperature.
PH CAL TO DO The Sensor Days value for the pH sensor is greater
than the Cal Reminder value
Calibrate the pH sensor.
TEMP CAL TO DO The Sensor Days value for the temperature sensor is
greater than the Cal Reminder value
Calibrate the temperature.
English 17
Table 4 Warning list for the sensor (continued)
Warning Description Resolution
PH MAINT RECD pH sensor maintenance is recommended
The slope is outside of the -50 to -61 mV/pH range
Clean the pH sensor, then repeat
the calibration with a fresh buffer
or sample, or replace the sensor.
PH MAINT RECD pH sensor maintenance is recommended
The offset is outside of ±45 mV but within ±60 mV
Clean the sensor and replace the
salt bridge and standard cell
solution and then repeat the
calibration, or replace the sensor.
T MAINT RECD The temperature offset is outside of ±3 °C but within
±5 °C
Calibrate the temperature.
Event log
The controller provides one event log for each sensor. The event log stores a variety of events that
occur on the devices such as calibrations done, calibration options changed, etc. A list of possible
events is shown in Table 5. The event log can be read out in a CSV format. For instructions on
downloading the logs, refer to the controller user manual.
Table 5 Event log
Event Description
Power On The power was turned on
Flash Failure The external flash has failed or is corrupted
1pointpHCalibration_Start Start of one-point sample calibration for pH
1pointpHCalibration_End End of one-point sample calibration for pH
2pointpHCalibration_Start Start of two-point sample calibration for pH
2pointpHCalibration_End End of two-point sample calibration for pH
1pointBufferpHCalibration_Start Start of one-point buffer calibration for pH
1pointBufferpHCalibration_End End of one-point buffer calibration for pH
2pointBufferpHCalibration_Start Start of two-point buffer calibration for pH
2pointBufferpHCalibration_End End of two-point buffer calibration for pH
TempCalibration_Start Start of temperature calibration
TempCalibration_End End of temperature calibration
pHCalSetDefault The pH calibration data was reset to the default
TempCalSetDefault The temperature calibration data was reset to the default
AllCalSetDefault All sensor calibration data was reset to the default
pHCalOptionChanged The pH calibration option was changed
TempCalOptionChanged The temperature calibration option was changed
SensorConfChanged The sensor configuration was changed
ResetpH CalHist The pH calibration history was reset
ResetTemp CalHist The temperature calibration history was reset
ResetAllSensorsCalHist All sensor calibration history was reset
ResetpHSensor The pH calibration data (sensor days, calibration history and calibration data)
was reset to the default
18 English
Table 5 Event log (continued)
Event Description
ResetTempSensor The temperature calibration data (sensor days, calibration history and
calibration data) was reset to the default
ResetAllSensors All sensor calibration data (sensor days, calibration history and calibration data)
was reset to the default
Replacement parts
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Table 6
Description Quantity Item no.
Sensor, pHD
2
1 9181500
Sensor, protective cap for pH
3
1 1000F3374-001
Standard cell solution 500 mL 25M1A1025-115
Gel powder (mixed with the standard cell solution for high temperature
applications)
2 grams 25M8A1002-101
Salt bridge (includes O-ring) 1 SB-R1SV
Buffer solution, pH 4 500 mL 2283449
Buffer solution, pH 7 500 mL 2283549
Buffer solution, pH 10 500 mL 2283649
2
Includes sealing hub for the pH flow cell.
3
Includes sponge to keep pH glass bulb wet during storage.
English 19
Technische Daten
Änderungen vorbehalten.
Technische Daten Details
Messbereich (pH) pH 2,5 bis 12,5
Messbereich (Temperatur) -5 bis 95 °C (23 bis 203 °F)
Auflösung pH 0,01 oder 0,1
Temperaturausgleich NTC-Thermistor, 300 Ohm
Stabilität (nur Analysegerät) 0,03 pH in 24 Stunden, nicht kumulativ
Empfindlichkeit Weniger als 0,005 pH
Max. Sondeneintauchtiefe/Druck 6,9 bar bei 105 °C (100 psi bei 221 °F)
Maximale Fließgeschwindigkeit 3 m (10 ft) pro Sekunde
Stromversorgung 5 VDC, 1 mA (Speisung erfolgt durch Controller)
Betriebstemperatur -5 bis 95 °C (23 bis 203 °F)
Kabel (Länge/Typ) Kabel 6 m (20 Fuß), 5-adrig (plus zwei isolierte Schirme) mit Mantel aus XLPE
(vernetztem Polyethylen); spezifiziert bis 150 °C (302 °F)
Max. Übertragungsdistanz 914 m (3000 Fuß)
Kalibriermethoden Erstmalige 2-Punkt-Kalibrierung mit 2 Puffern und der Möglichkeit, 1-Punkt-
oder 2-Punkt-Kalibrierung (Steilheit) mit Proben oder Puffern durchzuführen
Schnittstellen Modbus vom Gateway
Material Ryton
®
(PVDF) Körper, Salzbrücke in passendem Material mit Kynar
®
Dichtung, Glas-Prozesselektrode, Masseelektrode aus Titan und Viton
®
O-
Ring-Dichtungen
Allgemeine Angaben
Der Hersteller ist nicht verantwortlich für direkte, indirekte, versehentliche oder Folgeschäden, die
aus Fehlern oder Unterlassungen in diesem Handbuch entstanden. Der Hersteller behält sich
jederzeit und ohne vorherige Ankündigung oder Verpflichtung das Recht auf Verbesserungen an
diesem Handbuch und den hierin beschriebenen Produkten vor. Überarbeitete Ausgaben sind auf
der Hersteller-Webseite erhältlich.
Sicherheitshinweise
Bitte lesen Sie dieses Handbuch komplett durch, bevor Sie dieses Gerät auspacken, aufstellen oder
bedienen. Beachten Sie alle Gefahren- und Warnhinweise. Nichtbeachtung kann zu schweren
Verletzungen des Bedieners oder Schäden am Gerät führen.
Stellen Sie sicher, dass die Sicherheitseinrichtung dieses Messgerätes nicht beeinträchtigt wird.
Verwenden bzw. installieren Sie das Messsystem nur auf solche Art und Weise, wie sie in diesem
Handbuch beschrieben wird.
Verwendung der Gefahrenhinweise
G E F A H R
Weist auf eine potenzielle oder unmittelbare Gefahrensituation hin, deren Nichtbeachtung zu schweren
Verletzungen oder zum Tod führt.
W A R N U N G
Kennzeichnet eine mögliche oder drohende Gefahrensituation, die, wenn sie nicht vermieden wird, zum Tod oder
zu schweren Verletzungen führen kann.
20 Deutsch
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221
  • Page 222 222
  • Page 223 223
  • Page 224 224
  • Page 225 225
  • Page 226 226
  • Page 227 227
  • Page 228 228
  • Page 229 229
  • Page 230 230
  • Page 231 231
  • Page 232 232
  • Page 233 233
  • Page 234 234
  • Page 235 235
  • Page 236 236
  • Page 237 237
  • Page 238 238
  • Page 239 239
  • Page 240 240
  • Page 241 241
  • Page 242 242
  • Page 243 243
  • Page 244 244
  • Page 245 245
  • Page 246 246
  • Page 247 247
  • Page 248 248
  • Page 249 249
  • Page 250 250
  • Page 251 251
  • Page 252 252
  • Page 253 253
  • Page 254 254
  • Page 255 255
  • Page 256 256
  • Page 257 257
  • Page 258 258
  • Page 259 259
  • Page 260 260
  • Page 261 261
  • Page 262 262
  • Page 263 263
  • Page 264 264
  • Page 265 265
  • Page 266 266
  • Page 267 267
  • Page 268 268
  • Page 269 269
  • Page 270 270
  • Page 271 271
  • Page 272 272
  • Page 273 273
  • Page 274 274
  • Page 275 275
  • Page 276 276
  • Page 277 277
  • Page 278 278
  • Page 279 279
  • Page 280 280
  • Page 281 281
  • Page 282 282
  • Page 283 283
  • Page 284 284
  • Page 285 285
  • Page 286 286
  • Page 287 287
  • Page 288 288
  • Page 289 289
  • Page 290 290
  • Page 291 291
  • Page 292 292
  • Page 293 293
  • Page 294 294
  • Page 295 295
  • Page 296 296
  • Page 297 297
  • Page 298 298
  • Page 299 299
  • Page 300 300
  • Page 301 301
  • Page 302 302
  • Page 303 303
  • Page 304 304
  • Page 305 305
  • Page 306 306
  • Page 307 307
  • Page 308 308
  • Page 309 309
  • Page 310 310
  • Page 311 311
  • Page 312 312
  • Page 313 313
  • Page 314 314
  • Page 315 315
  • Page 316 316
  • Page 317 317
  • Page 318 318
  • Page 319 319
  • Page 320 320
  • Page 321 321
  • Page 322 322
  • Page 323 323
  • Page 324 324
  • Page 325 325
  • Page 326 326
  • Page 327 327
  • Page 328 328
  • Page 329 329
  • Page 330 330
  • Page 331 331
  • Page 332 332
  • Page 333 333
  • Page 334 334
  • Page 335 335
  • Page 336 336
  • Page 337 337
  • Page 338 338
  • Page 339 339
  • Page 340 340
  • Page 341 341
  • Page 342 342
  • Page 343 343
  • Page 344 344
  • Page 345 345
  • Page 346 346
  • Page 347 347
  • Page 348 348
  • Page 349 349
  • Page 350 350
  • Page 351 351
  • Page 352 352
  • Page 353 353
  • Page 354 354
  • Page 355 355
  • Page 356 356
  • Page 357 357
  • Page 358 358
  • Page 359 359
  • Page 360 360
  • Page 361 361
  • Page 362 362
  • Page 363 363
  • Page 364 364
  • Page 365 365
  • Page 366 366
  • Page 367 367
  • Page 368 368
  • Page 369 369
  • Page 370 370
  • Page 371 371
  • Page 372 372
  • Page 373 373
  • Page 374 374
  • Page 375 375
  • Page 376 376
  • Page 377 377
  • Page 378 378
  • Page 379 379
  • Page 380 380
  • Page 381 381
  • Page 382 382
  • Page 383 383
  • Page 384 384

Hach pHD Sensor Používateľská príručka

Typ
Používateľská príručka