4
It may be helpful to have your doctor note down the medication
dosage or an emergency telephone number on the card. Extra
cards are supplied with the device for this purpose.
7. Battery Indicator and Battery change
Low battery
When the batteries are approximately ¾ empty the battery symbol
AO
will flash as soon as the device is switched on (partly filled battery
displayed). Although the device will continue to measure reliably, you
should obtain replacement batteries.
Flat battery – replacement
When the batteries are flat, the battery symbol AO will flash as soon
as the device is switched on (flat battery displayed). You cannot
take any further measurements and must replace the batteries.
1. Open the battery compartment 7 at the back of the device by
pushing inwards at the two arrows and pulling out the battery
compartment cover.
2. Replace the batteries – ensure correct polarity as shown by the
symbols in the compartment.
Which batteries and which procedure?
Use 4 new, long-life 1.5 V, size AA alkaline batteries.
Do not use batteries beyond their date of expiry.
Remove batteries if the device is not going to be used for a
prolonged period.
Using rechargeable batteries
You can also operate this device using rechargeable batteries.
Only use «NiMH» type reusable batteries.
Batteries must be removed and recharged when the flat
battery symbol appears. They should not remain inside the
device as they may become damaged (total discharge as a
result of low use of the device, even when switched off).
Always remove the rechargeable batteries if you do not
intend to use the device for a week or more.
Batteries cannot be charged in the blood pressure monitor.
Recharge batteries in an external charger and observe the
information regarding charging, care and durability.
8. Using a Mains Adapter
You can operate this device using the Microlife mains adapter
(DC 6V, 600 mA).
Only use the
Microlife
mains adapter available as an original
accessory appropriate for your supply voltage.
Ensure that neither the mains adapter nor the cable are
damaged.
1. Plug the adapter cable into the mains adapter socket 5 in the
blood pressure monitor.
2. Plug the adapter plug into the wall socket.
When the mains adapter is connected, no battery current is consumed.
9. Error Messages
If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «ERR 3», is displayed.
* Please consult your doctor, if this or any other problem occurs
repeatedly.
If you think the results are unusual, please read through the
information in «Section 1.» carefully.
10.Safety, Care, Accuracy Test and Disposal
Safety and protection
• This device may only be used for the purposes described in
these instructions. The manufacturer cannot be held liable for
damage caused by incorrect application.
• This device comprises sensitive components and must be
treated with caution. Observe the storage and operating condi-
tions described in the «Technical Specifications» section.
Error
Description
Potential cause and remedy
«
ERR 1
» Signal too
weak
The pulse signals on the cuff are too weak.
Re-position the cuff and repeat the measure-
ment.*
«
ERR 2
» Error signal During the measurement, error signals were
detected by the cuff, caused for instance by
movement or muscle tension. Repeat the
measurement, keeping your arm still.
«
ERR 3
» No pressure
in the cuff
An adequate pressure cannot be generated
in the cuff. A leak may have occurred. Check
that the cuff is correctly connected and is not
too loose. Replace the batteries if neces-
sary. Repeat the measurement.
«
ERR 5
» Abnormal
result
The measuring signals are inaccurate and
no result can therefore be displayed. Read
through the checklist for performing reliable
measurements and then repeat the
measurement.*
«
HI
» Pulse or cuff
pressure
too high
The pressure in the cuff is too high (over 300
mmHg) OR the pulse is too high (over 200
beats per minute). Relax for 5 minutes and
repeat the measurement.*
«
LO
» Pulse too
low
The pulse is too low (less than 40 beats per
minute). Repeat the measurement.*