Tesla Smart Oximeter Používateľská príručka

Typ
Používateľská príručka
1
USER MANUAL
TESLA SMART
OXIMETER
EN
CZ
SK
PL
HU
SL
DE
2
THIS MEDICAL INSTRUMENT
MUST BE USED ACCORDING TO
INSTRUCTIONS TO ENSURE
ACCURATE READINGS
1. Power button
2. Display
3. SpO2 data
4. Bluetooth/Battery indicator
5. Pulse Rate data
6. Plethysmogram
7. Battery cover
4
5
3
2
16
7
3
1 SAFETY
1.1 Instructions for the Safe
Operation and Use of the
Tesla Smart Oximeter
• Do not attempt to service the Tesla
Smart Oximeter. Only qualified service
personnel should attempt any needed
internal servicing.
• Prolonged use or the patient’s condi-
tion may require changing the sensor
placement periodically. Change the
sensor placement and check sensor is
in contact with exposed skin, circulatory
status, and correct alignment at least
every 2 hours.
• Oxygen saturation (SpO2) measu-
rements may be adversely affected in
the presence of high ambient light. The
sensor area should be shielded with a
surgical towel, plaster, or bandage if
necessary.
• The following factors may cause
interference to the testing accuracy of
the Tesla Smart Oximeter:
4
• High-frequency electrosurgical
equipment.
• Placement of the sensor on an extre-
mity with a blood pressure impacted by
arterial catheter, or intravascular line.
• If the patient has severe hypotension,
vasoconstriction, severe anemia, or
hypothermia.
• If the patient is in cardiac arrest or
in shock.
• Fingernail polish or false fingernails
may cause inaccurate SpO2 readings.
• The device should be given at least
10 minutes to come to normal working
temperature if stored in a hot or cold
environment.
• The device is non-sterile and not
intended to be sterilized.
5
1.2 Warning
The MEDICAL ELECTRONIC
EQUIPMENT is suitable for home
healthcare environments:
• Although the medical electronic
equipment conforms to the intent of the
standard EN 60601-1-2 in relation to
electromagnetic compatibility, electrical
equipment may produce interference.
If interference is suspected, move the
equipment away from the sensitive
device.
• Portable and mobile RF commu-
nication equipment can affect this
instrument’s normal operation.
• Explosion hazard -Do not use the
Tesla Smart Oximeter in a flammable
atmosphere where concentrations
of flammable anesthetics or other
materials may occur.
• Do not throw batteries in fire as this
may cause them to explode.
• Do not attempt to recharge normal
dry-cell batteries, they may leak and
6
may cause a fire or even explode.
• Do not use the Tesla Smart Oximeter
in an MRI or CT environment.
• Do not modify this equipment without
the authorization of the manufacturer.
• If this equipment is modified, appro-
priate inspection and testing must be
conducted to ensure continued safe use
of equipment.
• Do not use near active high frequency
surgical equipment and the RF shielded
room of an medical electronic equip-
ment system for magnetic resonance
imaging, where the intensity of electro-
magnetic disturbances is high.
• Use of this equipment adjacent to or
stacked with other equipment should
be avoided, because it could result
in improper operation. If such use is
necessary, this equipment and the other
equipment should be observed to verify
that they are operating normally.
•Use of accessories, transducers, and
cables other than those specified or
7
provided by the manufacturer of this
equipment could result in increased
electromagnetic emissions or decreased
electromagnetic immunity of this equip-
ment and result in improper operation.
• Portable RF communications equi-
pment (including peripherals such as
antenna cables and external antennas)
should be used no closer than 30 cm
(12 inches) to any part of the Tesla Smart
Oximeter, including cables specified by
the manufacturer. Otherwise, it may re-
sult in a degradation of the performance
of this equipment.
• High-pressure sterilization cannot be
used on the device.
• IF ANY: a list of all cables and maxi-
mum lengths of cables (if applicable),
transducers and other accessories that
are replaceable by the responsible
organization and that are likely to affect
compliance with the requirements of
clause 7 (emissions) and clause 8 (im-
munity). accessories may be specified
8
either generically (e.g., shielded cable,
load impedance) or specifically (e.g., by
manufacturer and equipment or type
reference).
• IF ANY: the performance that was
determined to be essential performance
and a description of what the operator
can expect if the essential performance
is lost or degraded due to disturbances
(the defined term “essential performan-
ce” need not be used).
Cautions:
• Keep the operating environment
free of dust, vibrations, corrosive, or
flammable materials, and extremes of
temperature and humidity.
• Do not operate the unit, if it is damp or
wet because of condensation or spills.
Avoid using the equipment immediately
to a warm, humid location.
• Never use sharp or pointed objects to
operate the front-panel switches.
9
• The batteries must be taken out from
the battery compartment if the device
will not be used for a long time.
• The device shall only be used with the
battery cover closed.
• The batteries must be disposed of
properly, according to local regulation
aer their use.
• Keep the device away from children
and pets to avoid swallowing.
2 INTRODUCTION
2.1 Intended use
The Tesla Smart Oximeter is a non-inva-
sive device intended for spot checking
of functional oxygen saturation of
arterial hemoglobin (SpO2) and pulse
rate (PR). This portable device is suitable
for non-professional adult patients of
any color in clinical institutions and
home environments.
10
2.2 Brief Device Description
The Tesla Smart Oximeter is based
on digital technology. The device is
intended for non-invasive spot-check
measurement of functional oxygen
saturation of arterial hemoglobin
(SpO2). Advanced DSP algorithm can
minimize the influence of motion artifact
and improve measurement accuracy of
low perfusion.
The Tesla Smart Oximeter can be used
to measure human SpO2 and pulse
rate through the finger. The product is
suitable for private home-use, hospitals
(including clinical use in internist/
surgery, anesthesia, pediatrics etc.),
social medical organizations, physical
care in sports etc.
Note:
• DSP algorithm: Digital signal processor
algorithm.
• Low Perfusion: In physiology,
perfusion is the process of a body
11
delivering blood to a capillary bed in its
biological tissue. Under the condition
of low perfusion, the measurement of
non-invasive saturation of pulse-blood
oxygen is not accurate.
2.3 Product Features
• Lightweight for easy carrying and use.
• Manually adjust the direction of
interface.
• Color OLED display, simultaneous
display for testing value and plethys-
mography.
• Visual & Sound reminder function.
Real-time spot-checks.
• Wireless Bluetooth for data trans-
mission.
• Low Battery voltage indicator.
• Automatic switch-off.
• Includes two standard AAA 1.5V
alkaline batteries.
12
Note:
• Plethysmograph: is an instrument for
measuring changes in volume within an
organ or whole body (usually resulting
from fluctuations in the amount of
blood or air it contains).
CAUTION: The device must not be
used on children below 3 years of age,
as it cannot be guaranteed to provide
accurate results.
CAUTION: The Tesla Smart Oximeter
is intended only as an adjunct in patient
assessment. It must be used in conjun-
ction with other methods of assessing
clinical signs and symptoms.
CAUTION: A function tester cannot be
used to assess the accuracy of the Tesla
Smart Oximeter monitor or sensor.
Clinical testing is used to establish the
SpO2 accuracy. The measured arterial
SpO2 value (SpO2) of the sensor is
compared to arterial hemoglobin
oxygen (SaO2) value, determined from
13
blood samples with a laboratory CO-
oximeter. The accuracy of the sensors in
comparison to the CO-oximeter samples
measured over the SpO2 range of 70-100%.
Accuracy data is calculated using the root-
mean-square (Arms value) for all subjects.
Only about two--thirds of pulse oximeter
measurements can be expected to fall within
±Arms of the value measured by a CO-
oximeter. A pulse simulator shall be used to
assess pulse rate accuracy. The measured
pul-se rate is compared to the preset pulse
rate value in the simulator. Accuracy data is
calculated using the root-mean--square
(Arms value) for all subjects.
2.4 Expected Service Life
Is dependent on:
• The expected service life of the medical
electronic equipment;
• The expected service life of parts or acce-
ssories shipped with the medical electronic
equipment;
• Where the shelf life is less than the expect-
ed service life, the shelf life of parts or acce-
ssories shipped with the medical electronic
equipment;
14
3 INSTALLATION, SETUP
AND OPERATION
3.1 OLED display parameter setting
When the device is in the measuring
interface, press the direction button for 1
second in order to enter the menu page
(figure 3.1.1 and figure 3.1.2). There are
two sub-menus.
3.1.1 Reminder Setup
Press the direction button for 1 second
and enter the Reminder Setup. User
can adjust the setting through moving
the “*” symbol to the Sound Reminder,
Beep, Restore or Brightness.
• Sound reminder
Press the direction button for 1 second,
move the “*” symbol to the “Sound
Reminder”, long press the direction
button to turn the alarm sound on/
off. If the measured value exceeds the
maximum or minimum value of SPO2 or
PR, there will be an alarm sound, if this
function is turned on.
• Beep
15
Press the direction button for 1 second,
move the “*” symbol to the back of
“Beep”, long press the direction button
to turn the beep sound on/off. When
the beep is turned on, the sound
emitted during the test indicates the
pulse rate.
• Bluetooth
Press the direction button for 1 second,
move the “*” symbol to the back of
Bluetooth, long press the direction
button to turn it on/off.
• Restore
With the “*” symbol next to “Restore”,
long press the direction button. This
will change to “OK”, which restores the
device to factory settings.
• Brightness
With the “*” symbol next to “Brightness”,
long press the direction button to
change the Brightness value on a scale
from 1 to 5.
16
3.1.2 Limit Value Setting
With the “ * ” symbol next to “Reminder
Setup”, long press the direction button
until the “Limit Setup” menu appears.
You can then press the direction button
to select the items. Press the direction
button for 1 second to change to the
data you need.
On the “Limit Setup” menu page, with
the “*” symbol next to “+/-”, press the
direction button for 1 second to change
the “+” to “-” or change the “-” to “+”.
When “+” shows on the right side,
by pressing the direction button for
1 second and moving the “ * ” next to
the Spo2 Hi or PR Hi setting, you can
increase the value to a higher value
(until it reaches the maximum).
When “-” shows on the right side, by
press the direction button for 1 second
and moving the “ * ” next to the Spo2
Lo or PR Lo value setting, you can
reduce the value to a lower value (until it
reaches the minimum).
17
3.2 Operation
3.2.1 Installing batteries
Place two AAA batteries into the battery
compartment taking care to align the
polarities, then close the cover.
WARNING: Do not attempt to recharge
normal alkaline batteries, as they may
leak and may cause a fire or even
explode.
3.2.2 Turning on and applying the
Tesla Smart Oximeter
Put one of your fingers into the rubber
opening of the Tesla Smart Oximeter
with nail facing upward, then release the
clamp. Press the power button to turn the
Tesla Smart Oximeter on. The oximeter
will automatically turn off, if there is no
finger in the device for more than 16±2
seconds.
3.2.3 Read data from display screen
Display Description Show Note
OLED
Screen
The screen
display can
scroll
through
four
directions
with six
different
display
modes by
pressing the
power
button.
1. When the
battery power
is at its lowest
level, the
battery symbol
will be shown,
reminding
users to
replace the
batteries.
2. The plethy-
mogram can
been
considered
accurate if the
wave symbol
is fluctuating
regularly.
SpO2%
PR bpm
98
65
5.5
SpO2%
PR bpm
98
65
5.5
%SpO2
PR bpm
98
65
5.5
%SpO2
PR bpm
98
65
5.5
%SpO2
PR bpm
98
65
5.5
98
65
%SpO2
PR bpm
5.5
18
19
4 APP OPERATION
4.1 Preparation for use
• Tesla Smart is available for both iOS
and Android OS. Search the name ‚Tesla
Smart‘ in App Store or Android Market or
scan the QR-Code to download the App
and register or login your account.
• Make sure that the smartphone is co-
nnected to the 2.4Ghz network and have
Bluetooth & GPS location turned on.
4.2 Connect the device
• Open Tesla Smart app and click “+”
sign to Add new device
• Select Health care - Tesla Smart
Oximeter
20
• Please follow the instructions to add
the Tesla Smart Oximeter step by step
on your phone
4.3 Control device
• Control oximeter with APP
5 CLEANING AND DISINFECTION
5.1 Cleaning
Switch off the power and take out the
batteries before cleaning. Keep the
exterior surface of the device clean and
free of dust and dirt. Clean the exterior
surface (display screen included) of
the unit with a so dry cloth. Use 75%
medical alcohol to clean the surface by
applying a small amount with a dry cloth
to avoid the alcohol entering the device.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202

Tesla Smart Oximeter Používateľská príručka

Typ
Používateľská príručka