Tesla 1890095 Používateľská príručka

Kategória
Jednotky krvného tlaku
Typ
Používateľská príručka
1
USER MANUAL
TESLA SMART BLOOD
PRESSURE MONITOR
EN
CZ
SK
PL
HU
SL
DE
2
Please read the manual carefully before you use the unit and
keep the manual well after using.
SAFETY INFORMATION
• Your new digital blood pressure monitor uses the oscillometric
method of blood pressure measurement. This means the monitor
detects your blood´s movement through your brachial artery and
converts the movements into a digital Reading. An oscillometric
monitor does not need a stethoscope, so the monitor is simple
to use.
• Intelligent inflation will reduce the uncomfortable feeling by
incorrect inflation, and shorten the measurement time, prolong
the cuff ‘s usage lifetime.
• 2x199 sets memory function each measurement result will
be displayed on the screen and automatically stored. This unit
has blood classification index, could easy to check your blood
pressure.
• This automatic blood pressure monitor could measure the
systolic pressure diastolic pressure and pulse, the components
are included the body, cuff and printed instruction manual. This
unit is intended for adult use.
INTENDED USE
The automatic blood pressure monitor intended to measure
the systolic pressure, diastolic pressure, and pulse rate through
upper arm. They are expect used in the home, and hospital,
intended for over 12 years old adult using.
Contraindication
This product can‘t be used in patients who is with severe heart
insufficiency to avoid suffocation and death. This product is not
suitable for infants and children.
• To assure the correct use at the product, basic safety measures
should always be followed including the warning and the caution
listed in the instruction manual:
3
WARNING: This alert identities hazards that may cause
serious personal injury or death.
CAUTION: This alert identities hazards that may cause
minor personal injury. product damage, or property damage.
Those who have arrhythmia, diabetes, blood circulation or
apoplexy problem please use under the physician‘s instruction.
Contact your physician for specific information about your
blood pressure. Self-diagnosis and Treatment which use mea-
sured results may be dangerous. Follow The instructions of your
physician or licensed healthcare provider.
Please place on a high place where it can’t be touched by
children.
No modification of this equipment is allowed.
Do not modify This equipment without authorization of the
manufacturer.
If this equipment is modified, appropriate inspection and
testing must be conducted to ensure continued safe use of
equipment.
The cuff hose around the neck may cause the suffocation.
The swallowing of small part like packing bag, battery,
battery cover and so on may cause the suffocation.
Please don’t use a dilution agent, alcohol or, petrol to clean
the unit. Please don’t hit heavily or fall down the product from a
high place. Use The right cuff, otherwise, it cannot work.
Never leave any low battery in the battery compartment
since they may leak and cause damage To The unit.
Please take off the battery if you won’t use device in 3
months.
Replace The new batteries if the unit displays a low battery
symbol.
Do not mix the old and new batteries.
Please avoid using in a high radiant area to make your
measuring data correctly.
SAFETY INFORMATION
4
Do not use the equipment where flammable gas (such as
anesthetic gas, oxygen, or hydrogen) or flammable liquid (such
as alcohol) is present.
Do not touch the output of the AC adapter and the patient
simultaneously.
Do not touch the live end of battery and the patient simulta-
neously when changing the batteries.
WARNING
Do not dispose at electrical appliances as unsorted municipal
waste, use separate collection facilities. Contact your local
government for information regarding the collection systems
available. If electrical appliances are disposed at in landfills or
dumps, hazardous substances can leak into the groundwater and
get into the food chain, damaging your health and well-being.
1. Internally powered equipment;
2. Type BF applied part;
3. Protection against ingress of water or Particulate matter: IP21;
4. Not category AP /APG equipment;
5. Mode of operation: Intermittent operation;
The user must check that the equipment functions safely and
see that it is in proper working condition before being used.
CLASSIFICATION
5
Body
Display
Wide range Cuff
Cuff Connector
Air Socket for
Cuff Connector
User button
Memory button
Setting button
Start/Stop button
Type-C
PRODUCT STRUCTURE
6
The accessories cuff is L size, for upper-arm circumference 22-
42cm use. The cuff is treated as the applied part.
Insert the connector with cuff tube into the hole which is on the
left side of the device as picture. (Only provided cuff can be
used, cannot change to any other branded cuff.)
Remove the battery cover from the battery compartment and
insert the battery.
a) Remove the battery cover as picture showed.
b) Insert 4 AAA powerful batteries into the compartment and
ensure each battery is in the proper direction.
BATTERY INSTALLATION
7
Low battery and replacement
When power on an the low battery symbol “ ” will display
you must replace old batteries with new batteries, otherwise the
unit can‘t work.
Battery type and replacement
Please use 4pcs AAA identical 1.5V alkaline batteries. Do Not
use the batteries beyond their expiry date. Please remove the
batteries if you do not need to use device for a long Time.
WARNING
Dispose of the battery in accordance with all federal, state and
local laws. To avoid tire and explosion hazards, do Not burn or
incinerate the battery.
8
Adapter technical features:
Output voltage: Type-C 5V±5%
Output current: At least 600 mA
Note:
-When use USB adapter, the power of battery won‘t be consu-
med.
-When suddenly stop during measurement (like the plug off from
the outlet by carelessness), it must be reinserted the plug into the
unit, and restart the measurement.
User setting
Press button when power off, the screen will display or,
press button, it will be changed between “ ” and “ ”,
press button START/STOP when you confirm the user.
Unit setting
Press button SET when power on, the screen will display 0 or
0.0, then press button MEM, it will be changed between kPa or
mmHg. Press button SET when you confirm the unit.
SETTING MODE
9
PROPER USE OF THE UNIT
Measurement
Pre-measurement
• Relax for about five to ten minutes prior to the measurement
Avoid eating, drinking alcohol, smoking, exercising, and bathing
for 30 minutes before taking a measurement.
All these factors will influence the measurement result.
• Remove any garment that tits closely to your upper arm.
• Always measure on the same arm (normally left).
• Take measurement regularly at the same time of every day, as
blood pressure changes even during the day.
Common factors of wrong measurement
• All efforts by the patient to support their arm can increase
blood pressure.
• Make sure you are in a comfortable, relax position and do
not activate any of the muscles in the measurement arm during
measurement. Use a cushion for support if necessary.
• It the arm artery lies lower or higher than the heart, a false
reading will be obtained.
Note
• Only use clinically approved cuffs!
• A loose cuff or an exposed bladder causes a false reading.
• With repeated measurements, blood accumulates in the arm
which can lead to false reading,
Consecutive blood pressure measurements should be repeated
after a 1-minute pause or after the arm has been held up to allow
the accumulated blood to flow away.
10
Fitting the CUFF
1). Put the cuff on a table flatly with the velcro side
down. Pass The end of the cuff Through The metal
loop so that a circle is formed. The velcro closer will
now be facing outwards (ignore
this step if the cuff has already been prepared).
2). Push The cuff over the left upper arm so that the
tube points in the direction of the lower arm.
3). Wrap The cuff on the arm as illustrated. Make cer-
tain that the lower edge of the cuff lies approximately
2 To 3cm above the elbow and the rubber tube leaves
the cuff on the inner side of the arm.
4). Tighten The free end of the cuff and close the cuff
by affixing the velcro.
5). The cuff should be snug on your upper arm so that
you can fit 2 fingers between the cuff and your upper
arm. Any piece of clothing restricts the arm which must
be taken off.
6). Secure The cuff with the velcro closer in such a
way that it lies comfortably and not too tight. Lay your
arm on a table (palm upwards) so that the cuff is at the
same height as the heart. Do not bend the tube.
Note
If it is not possible to fit the cuff to your left arm, it can also be
placed on the right hand. However, all measurements should be
made using the same arm.
11
Measuring procedure:
After the cuff has been appropriately positioned, the measure-
ment can begin:
1). Press the START/STOP button, all symbols appear on the dis-
play, then the pump begins to inflate the cuff, the rising pressure
in the cuff is shown on the display.
2). Atter the suitable pressure has been reached, the pump stops,
and the pressure gradually falls. The cuff pressure is displayed. ln
base that the inflation is net sufficient, the device automatically
re—inflates to a higher pressure.
3) When the device detects the signal, the heart symbol on the
display starts to flash.
4). The measurement readings remain on the display until you
switch off the device. If no button is pressed for a period of 3
minutes, the device switches off itself to save the power.
Note
the symbol ” ” is displayed with the reading only if an irregu-
lar heartbeat is detected during the measurement.
Discontinuing a measurement
If it is necessary to interrupt a blood pressure measurement for
any reason (e.g., the patient feels unwell), the START/STOP
button can be pressed at any time. The device immediately
decreases the cuff pressure automatically.
Memory-recall of measurements
This blood pressure monitor automatically stores 2x199 sets
measurements value, the oldest record will be replaced by the
latest measurement value when more than 199 sets each user.
12
Read memory record
Press the button MEM when power Off, the latest 3 times
average value will be shown, press the button MEM again, the
Iast measurement value will be shown, as well as subsequent
measurements can be displayed one after the other by pressing
the button MEM each time.
Memory -clear of measurements
lf you are sure that you want to permanently remove all stored
memories. Press the button SET for 2 times until CL appears
when power off, press the START/STOP button, CL will flash for
4 times to clear all the memories. After this press button MEM,
and “no“ will be shown on the display which means that no
memory in-store.
1 Preparation for use
• Tesla Smart is available for both iOS and Android OS. Search
the name ‚Tesla Smart‘ in App Store or Android Market or scan
the QR-Code to download the App and register or login your
account.
• Make sure that the smartphone is connected to the 2.4Ghz
network and have Bluetooth & GPS location turned on.
APP OPERATION
13
2 Connect the device
• Open Tesla Smart app and click “+” sign to Add new device
• Select Health care - Tesla Smart Blood Pressure monitor
• Please follow the instructions to add the Tesla Smart Blood
Pressure monitor step by step on your phone
• Long press the „SET“ button of the blood pressure monitor, the
monitor will display the word „bt“ and enter the Bluetooth pairing
mode
3 Control the device
• Control blood pressure monitor with APP
4 Settings of users
• Device support Settings of 2 users.
• The ability to switch between user1/user2 and view the corre-
sponding user data are shown directly in the main APP interface.
• You can switch users both on the device and in the app
• When you are connected to the Tesla Smart app, after the
blood pressure monitor has finished measuring, the app switch
users automatically and the monitor will follow to synchronize
the app‘s user settings.
14
ABOUT BLOOD PRESSURE
Blood pressure is the pressure exerted the arteries,
The systolic blood pressure value represents the blood pressure
produced by contraction of the heart muscle.
The diastolic blood pressure value represents the blood pressure
produced by relaxation of the heart muscle.
• According to the blood pressure classification by the WHO/ISH.
• SYS lower than 100 mmHg (13.3kPa) is considered as hypotension.
15
EXCEPTIONAL SITUATION
Trouble removal
Error indicators
The following symbol will appear on the display when measuring
abnormal.
Please contact the distributor if you can‘t solve the problem, do not disassemble
the unit by yourself!
Symbol Cause Correction
Weak signal or pres-
sure change suddenly
Wrap the cuff properly.
Remeasure with correct way.
Weak signal or pres-
sure change suddenly
When near cell phone or other high
radiant devices the measurement will
be failed.
Keep quiet and no chatting when
measure.
It appears error during
the
process of inflating
Wrap the cuff properly.
Make sure that the air plug is properly
inserted in the unit.
Remeasure.
Abnormal blood
pressure
Repeat the measurement after relaxing
for 30 mins, if get unusual readings for 3
times please contact your doctor.
Low battery Replace all the worn batteries with new
ones.
Problems Check Cause and solutions
No power Wrap the cuff properly. Replace a new one
Check the polarity position Installation for proper placement
of the battery’s polarities
No inflation Whether the plug insert Insert into the air socket tightly
Whether move the arm when
inflating Change a cuff
Err and stop
working
Check if chatting when
measured Keep the body peaceful
Whether the cuff wrap too
loose Keep quiet when measuring
Cuff leak Whether the cuff broken Wrap the cuff tightly
Replace all the worn batteries
with new ones. Change a new cuff
16
CARE AND MAINTENANCE
Care for the main unit and blood pressure monitor cuff
• Keep the unit in the storage case when no use.
• Clean the unit with soft dry cloth.
• Do not use any abrasive or volatile cleaners.
• Never immerse the unit or any component in water.
• Make sure the monitor is off prior to cleaning, a mixture of
distilled water and 10 percent bleach could be used.
• Using a spray bottle, moisten a soft cloth towel with the bleach
or detergent mix until it is fully saturated. Squeeze any excess
moisture from the cloth to avoid any dripping or potential over
saturation of the cuff.
• Wipe all surfaces at the blood pressure monitor cuff thorou-
ghly, making sure to clean the inside and outside of the cuff. Be
cautious not to get any moisture in the main unit.
• Using a dry cloth, gently wipe away any excess moisture that
may remain on the blood pressure cuff. Lay the cuff flat in an
unrolled position and allow the cuff to air dry.
• Do not clean the body and cuff with naphtha, thinner or
gasoline, etc.
• Do not wet the cuff or attempt to clean the cuff with water
• Store the unit in a clean and dry location.
• Do not subject the unit to extreme hot or cold temperature,
humidity, and direct sunlight
• Remove the batteries it the unit will not be used in 3 months or
longer.
Maintenance
17
Description: Automatic upper arm blood pressure monitor
Display: LED digital display
Measuring principle: Oscillometric method
Measuring location: Upper arm
Measurement range:
Pressure: O~299mmHg
Pulse: 40~199 pulses/mm
Accuracy: Pressure: ±3mmHg
Pulse: ±5% of reading
LED indication: Pressure: 3 digits display of mmHg
Pulse: 3 digits display
Symbol: Memory/Heartbeat/low battery
Memory function: 2x199 sets memory of measurement values
Power source: 4pcs AAA alkaline battery/Type-C 5V
Automatic power off: in 3 minutes
Main unit weight: Approx.260g (batteries not included)
Main unit size: 138mm x 100mm x 62mm
Main unit lifetime: 10 000 times under normal use
Battery life: Could be used for 300 times for normal
condition
Accessories: Cuff, instruction manual
Operating environment: Temperature: 5°C~40°C
Humidity: 15%~93%RH
Air pressure: 86kPa~106kPa;
Storage environment: Air pressure: 86kPa~106kPa;
Temperature:-20°C~55°C; Humidity:
10%~93%RH;
avoid crash, sun burn or rain during trans-
portation
Expected service life 5 years
SPECIFICATION
18
EMC Declaration
IEC 60601 -1 -2:2014 Tesla Smart Blood Pressure Monitor: iden-
tification, marking and documents for Class B product
Instructions for use
The Tesla Smart Blood Pressure Monitor is suitable for home
healthcare environments and so on.
Warning: Don‘t hear active HF surgical equipment and the RF
shielded room of a system for magnetic resonance imaging,
where the intensity of EM disturbances is high.
Warning: Use of this equipment adjacent to or stacked with
other equipment should be avoided because it could result in
improper operation. If such use is necessary, this equipment and
the other equipment should be observed to verify that they are
operating normally,
Warning: Use of accessories, transducers and cables other
than those specified or provided by the manufacturer of this
equipment could result in increased electromagnetic emissions
or decreased electromagnetic immunity of this equipment and
result in improper operation.
Warning: Portable RF Communications equipment (including
peripherals such as antenna cables and external antennas)
should be used no closer than 30 cm (12 inches) to any part of
the blood pressure monitor, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this
equipment could result.
If any: a list of all cables and maximum lengths of cables (if appli-
cable), transducers and other accessories that are replaceable
by the responsible organization and that are likely to affect Tesla
Smart Blood Pressure Monitor with the requirements of Clause
7 (EMISSIONS) and Clause 8 (IMMUNITY). Accessories may
be specified generically (e.g., Shielded cable, load impedance)
or specifically (e.g., by manufacturer and equipment or type
reference].
If any: the performance of Tesla Smart Blood Pressure Monitor
that was determined to be essential performance and a descrip-
19
tion of what the operator can expect if the essential performance
is lost or degraded due to EM disturbances (the defined term
“essential performance „ need not be used).
Technical description
1. AII necessary instructions for maintaining basic safety and
essential performance regarding to electromagnetic disturban-
ces for the excepted service life.
2. Guidance and manufacturer’s declaration - electromagnetic
emissions and immunity.
Guidance and manufacturer’s declaration -
electromagnetic emissions
Emissions test Compliance
RF emissions CISPR 11 Group 1
RF emissions CISPR 11 Class B
Harmonic emissions IEC
61000-3-2
Class A
Voltage fluctuations/ flicker
emissions IEC 61000-3—3
Compliance
20
Guidance and manufacturer’s declaration - electromagnetic immunity
Immunity Test IEC 60601-1-2 Test level Compliance level
Electrostatic discharge
(ESD) IEC 61000-4-2
±8 kV contact, ±2 kV, ±4
kV, ±8 kV, ±15 kV air
±8 kV contact ±2
kV, ±4 kV, ±8 kV,
±15 kV air
Electrical fast transient/
burst IEC 61000-4-4
Power supply lines:
±2kV 100 KHZ repetition
frequency
Power supply
lines ±2kV 100
kHz repetition
frequency
Surge IEC 61000-4-5 line[s] to line ±1 kHz line(s) to line(s)
±1 kV
Voltage dips, short
interruptions and voltage
variations on power supply
input lines IEC 31000-4-11
0% 0.5 cycle at 0°, 45°,
90°, 135°, 180°, 225°, 270°
315° 0% 1 cycle and 70%
25/30 cycles single phase:
at 0 0% 300 cycles
0% 0.5 cycle at
0°, 45°, 90°, 135°,
180°, 225°, 270°
315° 0% 1 cycle
and 70% 25/30
cycles single
phase: at 0 0%
300 cycle
Power frequency magnetic
field IEC 61000-4-8
30 A/m, 50Hz/60Hz 30 A/m,
50Hz/60Hz
Conduced RF IEC 61000-
4-6
150 kHz 80 MHz 3Vrms
6Vrms (in ISM and amateur
[ln ISM and amateur radio
bands) 80% Am at 1kHz
150 kHz 80 MHz
3Vrms 6Vrms (in
ISM and amateur
[ln ISM and ama-
teur radio bands)
80% Am at 1kHz
Radiated RF IEC 61000-
4-3
10 V/m 80 MHz-2,7 GHz
80% AM at 1 kHz
10 V/m 80 MHz-
2,7 GHz 80% AM
at 1 kHz
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136

Tesla 1890095 Používateľská príručka

Kategória
Jednotky krvného tlaku
Typ
Používateľská príručka