Inogen GS-100 Používateľská príručka

Typ
Používateľská príručka
1
®
user manual
2
3
Contents
Chapter 1 1 Glossary of Symbols
Chapter 2 2 Warnings, Cautions and Contraindications
2 Contraindications
2 Warnings
3 Cautions
Chapter 3 4 Indications for Use
Chapter 4 4 User Interfaces, Controls and Serviceable Items
4 User Interfaces
5 User Controls
5 Serviceable Items
Chapter 5 6 Operating Instructions
Chapter 6 7 Audible and Visual Indicators
Chapter 7 10 Troubleshooting
Chapter 8 10 Cleaning, Care and Maintenance
10 Cleaning and Care
11 Routine Maintenance
11 Replacement Parts
11 Recommended Accessories
11 Service
12 Column Change Procedure
13 Storage
Chapter 9 13 Specifications
1
1. Glossary of Symbols
A warning indicates that
the personal safety of the
patient may be involved.
Disregarding a warning
could result in significant
injury
A caution indicates that
a precaution or service
procedure must be
followed. Disregarding a
caution could lead to a
minor injury or damage to
equipment
On/Off (Power)
Flow setting indicator
Blue Indicator Light
General alarm
indicator light
Follow instructions
for use
Type BF Applied Part
Class II Equipment
No Open Flames
No Smoking
No Oil or Grease
Do Not Disassemble
Electrical Safety Agency
Certificate
Magnetic Resonance
(MR) safe
Medical Device
U.S. Federal Regulation
Restricts this device to sale
by order of physician. May
also be applicable in other
countries
Do not dispose of in
unsorted municipal waste
AC Power
Keep Dry
Enclosure Ingress
Protection against
solid objects of 12.5mm
minimum diameter and
vertically falling drops of
water
Manufacturer
Complies with applicable
EU directives including
medical devices directive
Authorized representative
in the European
Community
Fragile
Humidity
Temperature
This Side Up
ONLY
ONLY
ONLY
WARNING
CAUTION
ONLY
ONLY
ONLY
ONLY
ONLY
As Packaged (Per Box)
Symbol Key
IP21
EC REP
2
Contraindications
This equipment is to be used as an oxygen supplement and is NOT
INTENDED to be life supporting or life sustaining.
Warnings
The device produces concentrated oxygen gas which accelerates
combustion. DO NOT USE THE DEVICE WHEN SMOKING OR NEAR OPEN
FLAME, MATCHES, PETROLEUM, OIL, GREASE, SOLVENTS, RADIANT HEATERS,
AEROSOLS etc. Use only water based lotions or salves that are concentrated
oxygen compatible during oxygen therapy.
Concentrated oxygen makes it easier for a fire to start and spread. Do not
leave the nasal cannula on upholstery or other fabric such as bedding or
personal clothing if the oxygen concentrator is turned on but not in use.
Concentrated oxygen will make the materials flammable. Turn off the
oxygen concentrator when not in use.
Do not use the oxygen concentrator in the presence of pollutants, smoke or
fumes, flammable anesthetics, cleaning agents or other chemical vapors.
This may internally contaminate the oxygen concentrator and degrade its
performance.
Do not use the oxygen concentrator if either the plug or the power cord is
damaged to prevent accidental electrical shock.
Do not submerge the oxygen concentrator in liquids, expose to liquids, or
otherwise allow liquids to enter into the case, this may lead to electrical
shock and/or damage. If the oxygen concentrator is exposed to liquids turn
it off and unplug from electrical outlet before attempting to clean and dry
the liquid spill.
When using a compatible humidifier do not refill humidifier while
attached to the oxygen concentrator. Remove the humidifier from the
oxygen concentrator prior to refilling to prevent accidental spills on the
concentrator and prevent electrical shock hazard.
Do not use cleaning agents other than those specified in this manual.
Always unplug the oxygen concentrator before performing any cleaning.
Do not use alcohol, isopropyl alcohol, ethylene chloride, or petroleum based
cleaners. This may impair the proper function and/or increase the risk of
fire and burns.
Do not disassemble the oxygen concentrator or attempt any maintenance
other than tasks described in this user manual; disassembly creates a
hazard of electrical shock and will void the warranty. Servicing of the
concentrator is to be performed by qualified and trained personnel only.
Do not use any columns other than those specified in this user manual.
The use of non-specified columns may create a safety hazard and/or impair
equipment performance and will void the warranty.
2. Warnings, Cautions and Contraindications
3
To avoid danger of choking and strangulation hazard, keep tubing away
from children and pets.
If you begin to feel ill or are experiencing discomfort while using the oxygen
concentrator, consult your health care provider immediately.
Cautions
USA Federal law restricts this device to sale or rental by or on the order of
a physician, or any other practitioner licensed by the State in which he/
she practices to use or order the use of this device; may also be applicable
in other countries. Under certain circumstances, the use of non-prescribed
oxygen therapy can be hazardous.
Availability of an alternate source of oxygen is recommended in case of
power outage or mechanical failure. Consult your equipment provider for
type of back-up system recommended.
Additional monitoring or attention may be required for patients using this
device who are unable to hear or see alarms or communicate discomfort. If
the patient shows any signs of discomfort, a physician should be consulted
immediately.
Accessories not specified for use with the oxygen concentrator may
impair performance. Always use accessories according to manufacturer’s
instructions.
Nasal cannula should be rated 5 liters per minute to ensure proper patient
usage and oxygen delivery.
Replace the nasal cannula on a regular basis. Check with your equipment
provider or health care professional to determine how often the cannula
should be replaced.
It is recommended the oxygen concentrator accessory tubing and cannula
include a means to reduce the propagation of fire.
Do not operate the concentrator without the intake filter and particle filter
in place. Particles drawn into the oxygen concentrator may damage the
equipment.
Refer to the environmental range specifications for proper storage and use
conditions. Temperatures in excess of the indicated range may cause device
malfunction.
Do not obstruct air intake or exhaust when operating the oxygen
concentrator. Blockage of air circulation or proximity to a heat source may
lead to internal heat buildup and shutdown or damage. Always maintain a
minimum of six inches of clearance.
Do not place anything in the power supply port other than the supplied
wall cord. Avoid the use of electrical extension cords with the concentrator.
If an extension cord must be used, use an extension cord that has an
Underwriters Laboratory (UL) Mark and a minimum wire thickness of 18
gauge. Do not connect any other device to the same extension cord.
Do not sit or stand on the concentrator, doing so can be hazardous.
4
In order to ensure a safe installation and operation of the Inogen At Home
Oxygen Concentrator Model GS-100, read and understand this entire manual
before using the device.
The Inogen At Home Oxygen Concentrator has a 5 year expected life except
1 year expected life for user serviceable sieve bed columns.
3. Indications for Use
The Inogen At Home Oxygen Concentrator is used on a prescriptive basis by
patients requiring supplemental oxygen. It supplies a high concentration
of oxygen and is used with a nasal cannula to channel oxygen from the
concentrator to the patient. The Inogen At Home Oxygen Concentrator may be
used in a home or institution.
4. User Interfaces, Controls and Serviceable Items
On/Off Button
Flow Setting Indicator
Blue Indicator
Light
General Caution Indicator
Flow
Control
User Interfaces
General Caution Indicator Light (Low, Medium Priority Alarm)
A yellow light indicates either a change in operating status or
a condition that may need response. A flashing light is higher
priority than non-flashing.
Blue Indicator Light (Change Column Soon)
Column maintenance is required within 30 days.
Contact your equipment provider to arrange service.
5
User Interfaces
Audible Indicators
An audible alarm (beep) indicates either a change in operating
status or a condition that may need response (alarm). More
frequent beeps indicate higher priority conditions.
Flow Setting Indicator
A green indicator light indicates the selected flow setting.
User Controls
ON / OFF Button
Press once to turn “ON”; Press and hold for one second to turn
“OFF”.
Flow Setting Control Buttons
Use the – or + flow setting control buttons to select the setting
as shown on the display. There are five settings, from 1 to 5.
Serviceable Items
Particle Filter
The filter must be in place at the intake end of the
concentrator during operation.
Air Inlet Filter
The filter must be in place at the top of the
concentrator during operation.
Cannula Nozzle Fitting
The nasal cannula connects to this nozzle
for output of oxygen.
Nasal Cannula
Nasal cannulas should be replaced on a regular basis, consult
with your physician or equipment provider, or cannula
manufacturer’s instructions.
Nozzle
Fitting
6
5. Operating Instructions
1. Place the concentrator in a well-ventilated location; air
intake and exhaust must have clear access. Ensure
that the concentrator is at least 6 inches (15 cm.)
away from walls, furniture and curtains that could
impede adequate airflow to the device.
1. Ensure particle filter is in place.
2. Ensure intake filter is in place.
3. Follow instructions A or B below:
A. If you are not using a humidifier bottle, connect
your nasal cannula tubing to the nozzle fitting.
Nozzle fitting is located on the top of the
concentrator. See illustrations 4A1 and 4A2.
B. If you are using a humidifier bottle, follow the
manufacturer’s instructions. Place the humidifier
bottle in the designated holder.
See illustrations 4B1 and 4B2.
5. Attach the power cord to the back of the concentrator
and plug the other end into an electrical outlet. Do not
position the concentrator so that it is difficult to
disconnect the power cord. Turn on the concentrator by
pressing the ON/OFF button .
See illustration 5. A single short beep will sound and all
indicator lights will illuminate for a few seconds.
6. Use the or buttons to adjust the device to the
prescribed setting. There are five flow settings, from
1 liter per minute to 5 liters per minute. The current
setting can be viewed on the display. The green LED will
illuminate once the set flow rate of oxygen is achieved.
You may begin breathing from the device; the required
oxygen concentration is normally reached within five
minutes after device is powered on.
7. Ensure that the tubing is not kinked or pinched in any
way and that oxygen is flowing through the nasal
cannula. If oxygen is not flowing, the green LED will not
be illuminated. Refer to the troubleshooting guide
section of this manual.
8. Adjust the nasal cannula so that it is properly aligned
on your face or as directed by your healthcare provider.
9. Turn off the concentrator by pressing the ON/OFF
button . Turn off the concentrator when not in use.
4A2
4A1
Nozzle
Fitting
4B1
4B2
5
7
6. Audible and Visual Indicators
The concentrator has an audible alarm and three visual indicators
(green, yellow and blue).
Notifications
The concentrator monitors various parameters during operation and utilizes
an intelligent alarm system to indicate a malfunction of the concentrator.
Mathematical algorithms and time delays are used to reduce the probability
of false alarms while still ensuring proper notification of an alarm condition
If multiple alarm conditions are detected, the highest priority alarm will be
displayed.
To insure that audible notifications may be heard, the operators position must
be determined to suit the surrounding noise level.
Note that failure to respond to the cause of an alarm condition for low, medium
and high priority alarms potentially will result in discomfort or reversible minor
injury only and develop within a period of time sufficient to switch to a backup
source of oxygen.
The following notification messages are accompanied by an audible and/or
visual indicator.
Yellow
Indicator Light
Blue
Indicator Light
Flow Setting
Indicator
Condition/Action/Explanation
Current flow rate of oxygen being displayed at a
setting from 1-5 liters per minute
Flow status indicator is green
Column maintenance is required within 30 days.
Contact your equipment provider to arrange service.
Indicator
Flow rate on LED
Display
Flow rate indicator
Blue Indicator Light
8
Low Priority Alarms
The following low priority alarm messages are accompanied by a double beep
and a solid yellow light.
Concentrator is producing oxygen at a slightly
decreased level. Contact your equipment
provider to arrange service.
The concentrator is operating to specification
but requires service at the earliest convenience.
Contact your equipment provider to arrange
service.
One of the concentrator’s sensors has
malfunctioned. If the condition persists, contact
your equipment provider to arrange service.
The concentrator is producing oxygen at a
slightly decreased rate. If condition persists,
contact your equipment provider to arrange
service.
Low Oxygen
Service Needed
Sensor Fail
Low Flow
4
3
2
Action/ ExplanationCondition
Indicator
Proper oxygen delivery has not been detected.
Check for kinked tubing. If condition persists,
switch to a backup oxygen source and contact
your equipment provider to arrange service.
Flow Error
4
Action/ Explanation
The concentrator is producing oxygen at a
low concentration. Switch to a backup oxygen
source and contact your equipment provider to
arrange service.
Condition
Oxygen Error
Indicator
5
Medium Priority Alarms
The following medium priority alarm messages are accompanied by a triple beep,
repeated every 25 seconds, and a flashing yellow light.
Flow setting flashing
9
High Priority Alarms
The following high priority alarm messages are accompanied by a five beep
pattern, repeated every 10 seconds and a flashing yellow light warning
indicating immediate response by the operator is required.
Action/ Explanation
Smoke has been detected inside the device and
device is shutting down. Allow the concentrator
to air out in a smoke-free environment and
restart. If condition persists, switch to a backup
source of oxygen and contact your equipment
provider.
Condition
Fire Hazard
Indicator
5
Concentrator temperature is too cold for
operation. Allow the concentrator to warm
up in a room temperature environment for 10
minutes and restart. If condition persists, switch
to a backup source of oxygen and contact your
equipment provider.
Switch to a backup source of oxygen and
contact your equipment provider.
Concentrator lost power during operation.
Try unplugging the power cord and plugging
it back in. If alarm does not reset, press and
hold the power button down for 2 seconds. If
condition persists, switch to a backup source of
oxygen and contact your equipment provider.
Concentrator temperature is too high for
operation. Ensure air intake and outlet vents
have clear access and particle filters are clean.
Allow the concentrator to cool for 10 minutes
and restart. If condition persists, switch to a
backup source of oxygen and contact your
equipment provider.
System Cold
System Error
Power Error
System Hot
3
2
Sound indicator
only
4
NOTE When power is lost during operation, the buzzer will alarm
continuously.
10
8. Cleaning, Care and Maintenance
Cleaning and Care
Periodically you may clean the outside case using a cloth dampened with a mild
liquid detergent (such as Dawn™) and water.
Follow manufacturer’s instructions on cleaning and care of concentrator
accessories; clean or replace these accessories according to your healthcare
provider or respective manufacturer’s instructions for use.
Filter Cleaning and Replacement
The particle filter must be cleaned weekly to ensure the ease of air flow.
Remove filter from the side of the device.
7. Troubleshooting
Check power cord for
proper connection
Contact your
equipment provider
Press On/Off button to
power concentrator
Check cannula and its
connection
Inspect and replace
the items if necessary
Power cord is not
connected properly
Malfunction
Concentrator is not
powered on
Cannula is not
connected properly or is
kinked or obstructed
The oxygen tubing or
cannula is faulty
Concentrator does not
power on when On/Off
button is pressed
No oxygen
Recommended Solution
Refer to Chapter 4
Possible Cause
Refer to Chapter 4
Problem
Any problem accompanied
by information on
concentrator display,
visual and/or audible
indicators
Contact your home care provider if you need assistance with the device.
11
Clean the particle filter with a mild liquid detergent (such
as DawnTM) and water; rinse in water and dry before reuse.
Ensure the filter is completely dry before reinstalling into the
concentrator.
Air Inlet Filter
Replace when filter is dirty according to visual indicator.
The air inlet filter may be replaced by the equipment provider
or by the user.
Output Filter
The output filter is intended to protect the user from
inhalation of small particles in the product gas flow. The
device includes an output filter conveniently located behind
the removable cannula nozzle fitting. Inogen requires that this
filter be replaced between patients.
The output filter may be replaced by the equipment provider or by the user.
Routine Maintenance
No special maintenance, other than weekly particle filter cleaning, needs to be
carried out by the user.
Replacement Parts
Contact your home care provider if you have any questions about equipment.
Use only the following replacement parts:
Inogen At Home Particle Filter (RP-400)
Inogen At Home Output Filter Replacement Kit (RP-107)
Inogen At Home Air Inlet Filter (RP-401)
Inogen At Home Column Pair (RP-402)
AC Power Cord (RP-109)
Recommended Accessories
Humidifier Connector Tube (Salter Labs #S0-676)
Humidifier Bottle (Salter Labs #7600)
Nasal Cannula (Salter Labs #16SOFT)
Service
The concentrator is specifically designed to minimize routine preventative
maintenance.
Air Inlet
Filter
Particle Filter
12
For assistance, if needed, in setting up, using, maintaining, or reporting
unexpected operation or events, contact your equipment provider, or the
manufacturer.
Column Change Procedure
There are two columns (metal tubes) on the concentrator located behind the
particle filter. These two columns will be removed and replaced when service
is required. Ensure adequate time to complete all steps without disruption is
alloted when performing this maintenance.
1. Turn off the concentrator by pressing the power button to shut down the
device.
2. Unplug the concentrator from the electrical outlet.
3. Lay down the concentrator on its side.
4. Press the two latch buttons on the sides of the device to remove the cover
from the bottom of the concentrator in order to access the two columns.
5. Press the latch button with thumb or finger and pull the column out of the
concentrator using the pull ring attached to the bottom of the column cap.
6. Remove the column completely from the concentrator.
7. Repeat steps 5-6 to remove the other column.
Installation of the new columns (metal tubes):
8. Remove the upper and lower dust caps from each column. There should be a
total of two dust caps removed from each column. Make sure there is no dust or
debris where the dust caps were located.
9. Insert the new column into the concentrator . Do not leave the column
exposed after the dust caps are removed; it should be inserted into the
concentrator as soon as the dust caps have been removed, to minimize
environmental exposure.
10. The spring loaded latch button should fully return to the closed position
when the column is completely inserted.
11. Repeat steps 9-10 to install the other column.
12. Replace the lower cover and return the concentrator to its upright position.
13. Connect the power supply cord from the concentrator and plug into an
electrical outlet. Do Not Power On the concentrator.
14. Press and hold the plus button for ten seconds. Flow indicators 1, 3 & 5
will be illuminated and column will be reset. Release the plus button.
15. Press the power button to turn on the concentrator and use normally.
NOTE Column change instructions are only to be used when maintenance is
required. Columns should only be removed during this maintenance
procedure.
13
Storage
When not in use, store indoors away from excessive moisture and
temperatures. Storage conditions outside of range specified may result in
damage and device malfunction.
Disposal of Equipment and Accessories
Follow your local governing ordinances for disposal and recycling of the
concentrator and accessories.
Dimensions: 16.5”H x 13”W x 7”D
Weight: 18 lbs
Warm-Up Time: Less than 5 minutes
Oxygen Concentration: 90 +6%/-3% at all settings
AC Power: 100-240VAC, 275 W Max, 50-60Hz
Environmental Ranges
for Use: Temperature: 41 to 104°F (5 to 40˚C)
Humidity: 15 to 95%, non-condensing
Altitude: 0 to 10,000 ft (0 to 3048 m)
Environmental Ranges for
Shipping & Storage: Temperature: -13 to 158°F (-25 to 70˚C)
Humidity: 0 to 93% non-condensing
Altitude: n/a
Maximum Outlet Pressure: 19.6-40.6 psia
(135-280 KPa Absolute)@ 20˚C
Flow Control Settings: 5 Settings: 1 to 5 liters per minute
9. Specifications, Inogen At Home, Model GS-100
Standards Compliance
This device is designed to conform to the following standards:
EN ISO 8359:2012, Oxygen Concentrators for Medical Use – Safety
Requirements (ISO 8359:1996, Second edition, Amendment 1)
ASTM F 1464:2005, Oxygen Concentrators for Domiciliary Use
IEC 60601-1:Edition 3.1: 2012, Medical Electrical Equipment
-- Part 1: General requirements for basic safety and essential performance
Classification:
IEC Class II Equipment
Type BF Applied Part
IP21 Enclosure Ingress Protection against solid objects of 12.5mm minimum
diameter and vertically falling drops of water
Not suitable for use in the presence of a flammable anesthetic mixture
with air or with oxygen or nitrous oxide
Continuous Operation
14
Portable and mobile RF communications equipment should be
used no closer to any part of the device, including cables, than
the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance:
d=1.2√P 150 kHz to 80 MHz
d=1.2√P 80 MHz to 800 MHz
d=2.3√P 800 MHz to 2.5 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation distance
in meters (m).
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey a, should be less than the
compliance level in each frequency rangeb.
Interference may occur in the vicinity of equipment marked
with the following symbol:
3 Vrms
150 kHz
to 80 MHz
3V/m
80 MHz
to 2.5 GHz
3 Vrms
3V/m
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Compliance
Level
IEC 60601
Test Level
Immunity Test Electromagnetic
Environment - Guidance
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity:
The Concentrator is intended for use in the electromagnetic environment specified below. The user
of the Concentrator should make sure it is used in such an environment.
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30 %.
Mains power quality should be that of a typical commercial
or hospital environment.
Mains power quality should be that of a typical commercial
or hospital environment.
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
EC 61000-4-4
Surge
IEC 61000-4-5
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
± 2 kV for
power supply
lines
± 1 kV for
input/output
lines
± 2 kV for
power supply
lines
± 1 kV for
input/output
lines
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
± 2 kV line(s)
to earth
Mains power quality should be that of a typical commercial
or hospital environment. If the user of the [ME EQUIPMENT or
ME SYSTEM] requires continued operation during power mains
interruptions, it is recommended that the
[ME EQUIPMENT or ME SYSTEM] be powered from
an uninterrupted power supply or a battery.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical hospital or home
environment.
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
Power
frequency
(50/60 Hz)
magnetic
field
IEC 61000-4-8
<5% UT (>95%
dip in UT) for
0.5 cycle
40% UT (60%
dip in UT) for
5 cycles
70% UT (30%
dip in UT) for 25
cycles
<5% UT (>95%
dip in UT) for
5 sec
<5% UT (>95%
dip in UT) for
0.5 cycle
40% UT (60%
dip in UT) for
5 cycles
70% UT (30%
dip in UT) for 25
cycles
<5% UT (>95%
dip in UT) for
5 sec
3 A/m 3 A/m
15
NOTE At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
NOTE UT is the a.c. main voltage prior to application of the test level.
a: Field strength from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the concentrator is used exceeds the
applicable RF compliance level above, the concentrator should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the device.
b: Over the frequency range 150 kHz to 80 MHz, the field strengths should be less than 3V/m.
Recommended Separation Distances between Portable and Mobile RF Communications
Equipment and This Device:
This concentrator is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The user of the concentrator can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and this concentrator as
recommended below, according to the maximum output power of the communications equipment.
150 kHz to 80 MHz
d=1.2√P
0.12
0.38
1.2
3.8
12
0.12
0.38
1.2
3.8
12
0.23
0.73
2.3
7.3
23
0.01
0.1
1
10
100
80 MHz to 800 MHz
d=1.2√P 800 MHz to 2.5 GHz
d=2.3√P
Separation Distance According to Frequency of Transmitter (M)Rated Maximum Power
Output of Transmitter (W)
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maxi-
mum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The concentrator is intended for use in the electromagnetic environment specified below. The user of the
concentrator should assure that it is used in such an environment.
NOTE At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE The guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
Emissions Test Compliance Electromagnetic Environment - Guidance
RF emissions
CISPR 11 Group 1 The concentrator uses RF energy only for its internal function.
Therefore its RF emissions are very low and not likely to cause any
interference in nearby equipment.
RF emissions
CISPR 11 Class B The concentrator is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic Emissions
IEC 61000-3-2 Class A
Voltage fluctuations
/ flicker emissions
IEC 61000-3-3
Complies
16
17
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147

Inogen GS-100 Používateľská príručka

Typ
Používateľská príručka