BDS_sro_NAMO_DM1X_27.11.2019
English
www.bdsensors.com
WARNING! Install the device only in depressurized and currentless state!
WARNING! Use a suitable seal, corresponding to the medium and the pressure input.
DO NOT USE ANY ADDITIONAL SEALING MATERIALS; LINKE YARN, HEMP OR TEFLON TAPE
(for connection acc. DIN 3852)!
Connection acc. to: NPT
DIN 3852 EN 387 connections UNF
G1/2": G1/4”: 1/4" NPT: 7/16“-20:
ca. 10 Nm ca. 20Nm ca. 30 Nm ca. 20 Nm
The indicated tightening torques must not be exceeded!
Návod k montáži /
Mounting instructions
Digitální manometr / Digital pressure gauge
DM10, DM17
Centrála /
Headquarters
BD SENSORS s.r.o.
Hradišťská 817
CZ - 687 08 Buchlovice
Česká republika / Czech Republic
Tel.: +420 572 / 411 011
Fax: +420 572 / 411 497
Západní Evropa /
Western Europe Headquarters
BD SENSORS GmbH
BD-Sensors-Str. 1
D - 95199 Thierstein
Německo /
Germany
Tel.: +49 (0) 92 35 / 98 11-0
Fax: +49 (0) 92 35 / 98 11-11
Rusko /
Russia
BD SENSORS RUS
39a, Varshavskoe shosse
RU - Moscow 117105
Rusko /
Russia
Tel.: +7 (0) 9 59 81 / 09 63
Fax: +7 (0) 9 57 95 / 07 21
Tento návod k montáži je výtah z podrobného
návodu k obsluze. Ten si prosím stáhněte
z našich internetových stránek, pokud nejste
s výrobkem dobře seznámeni.
These mounting instructions are an excerpt from the complete
operating manual. It may be downloaded from our homepage,
if you are not familiar with the device.
www.bdsensors.cz
www.bdsensors.com
– Technické změny vyhrazeny –
– Technical modifications reserved –
WARNING! In order to avoid hazards to operators and damages
to the device, the following instructions have to be performed by
qualified technical personnel.
WARNING! Adhere to the safety and operating instructions stated in
the operation manual. Effective regulations on occupational safety,
accident prevention as well as national installation standards and
approved engineering techniques must in addition be complied with.
WARNING! Before the installation: make sure that the inside of the
threaded neck of the closed valve is clean. If need, remove the residual
liquid in a proper way. Do not mount the device against the liquid
inside the thread connection.
Limitation of liability
If the instructions in the operating manual are not adhered
to or if the device is inappropriately used, modified or
damaged, liability is not assumed and warranty claims will
be excluded.
Intended use
Ensure that the medium is compatible with the media-
wetted parts and that the device is suitable for the
application without restrictions. The technical data listed in
the current data sheet is binding and must definitely be
observed.
Product identification
Changing the batteries
To change the battery go ahead as follows:
- Remove the cap and change the battery.
- Lock the device after that properly.
Structure of the menu system
Mounting
Display and operating module
Structure of the menu system
Positioning of the display module
The display module of the pressure gauge is rotatable so that
clear readability is guaranteed even on unusual installation
positions.
Display
Select / ON -button
Function / OFF -button
:
- switch-on of the device - switch-off of the device
- choosing of the pressure units - scrolling in the menu system
- calibration of starting point
- configuration of the switch-off automatic
1 MIN
The minimum pressure during the measurement process will be shown in the display.
To delete the stored value push the Select/ON-button. It appears in the display, the value has
been recessed.
2 MAX
The maximum pressure during the measurement process will be shown in the display.
To delete the stored value push the Select/ON-button. It appears in the display, the value has
been recessed.
3 UNIT
Setting of the pressure unit
Possible units are: bar, mbar, psi, MPa, mH
2
O
The desired unit may be selected and activated with the Select/ON-button.
Depending on the nominal pressure range and the accuracy of the device, perhaps not all available
units could be used.
If you detect a shifting of the measured value deviating from the offset, the display can be re-calibrated via
pushing the Select/ON-button. The display shows , the value has been recessed. Differs the zero
point from the ambient pressure, it is a pressure reference necessary.
5 SW OFF
Configuration of the switch
The desired switch-off automatic may be selected through the Select/ON-button.
Allocation of the programmable values:
"0": switch-off automatic is not active
"1" – "5": switch-off automatic after 1 up to 5 minutes
The menu system will be leaved automatically after 10 seconds, the last setted value has been saved. If you scroll all menu
points you will leave the menu system after
SW OFF
.
After configuring the unit, the conversion of the pressure range into the new unit will only occur after leaving the menu
system. Depending on the pressure range, probably not all available units can be used.
ordering code
type designation input serial
number
Take note that no inadmissibly high mechanical
stress occurs at the pressure port, since this
may cause a shifting of the characteristic curve
or to the demage. This is especially important
for very small pressure ranges as well as for
devices with a pressure port made of plastic.
For more information, please read the